摘要
Background. Bleeding from coagulopathic hemodialysis puncture sites can contribute to anemia in dialysis patients, and current compressive dressings may contribute to graft thrombosis. We studied the safety and efficacy of a new chitosan-based bandage with an active clotting surface and compared its time to hemostasis and compression strap usage in dialysis access puncture wounds with that of conventional gauze dressings. Methods. Fifty patients received both the chitosan-based and conventional gauze dressings in random order on 2 successive visits. Time to hemostasis and compression strap usage were compared between the visits. Time to hemostasis was analyzed using the binary response variable at 2 and 4 minutes. A compression strap was used if dressing application was unsuccessful at 4 minutes. Covariates included coagulation state as measured by laboratory analysis and anticoagulation therapy. Results. Hemostasis was achieved by 2 minutes in 30 % of the chitosan-based and 38 % of the conventional dressings (p = 0. 608) and by 4 minutes in 86 % of the chitosan-based and 72 % of the conventional dressings (p = 0. 040). Compression strap usage was reduced by 50 % in the chitosan-based group compared to the conventional group (7 vs. 14 patients; p = 0. 052). No adverse events were reported with either dressing. Conclusions. The chitosan-based bandage from HemCon is a safe and effective hemostatic agent to reduce prolonged post-hemodialysis puncture site bleeding and may reduce the use of occlusive compression straps.
引用
收藏
页码:672 / +
页数:7
相关论文
共 12 条
[1]
Use of chitosan bandage to prevent fatal infections developing from highly contaminated wounds in mice [J].
Burkatovskaya, Marina ;
Tegos, George P. ;
Swietlik, Emilia ;
Demidova, Tatiana N. ;
Castano, Ana P. ;
Hamblin, Michael R. .
BIOMATERIALS, 2006, 27 (22) :4157-4164
[3]
Feldman HI, 1996, J AM SOC NEPHROL, V7, P523
[4]
Comparison of structural and hemostatic properties of the poly-N-acetyl glucosamine Syvek Patch with products containing chitosan [J].
Fischer, TH ;
Connolly, R ;
Thatte, HS ;
Schwaitzberg, SS .
MICROSCOPY RESEARCH AND TECHNIQUE, 2004, 63 (03) :168-174
[5]
GUSTAFSON S, UNPUB PREHOSPITAL EM
[6]
COMPARING CLASSIFICATION OF SUBJECTS BY 2 INDEPENDENT JUDGES [J].
MAXWELL, AE .
BRITISH JOURNAL OF PSYCHIATRY, 1970, 116 (535) :651-&
[7]
NOTE ON THE SAMPLING ERROR OF THE DIFFERENCE BETWEEN CORRELATED PROPORTIONS OR PERCENTAGES [J].
McNemar, Quinn .
PSYCHOMETRIKA, 1947, 12 (02) :153-157
[8]
Stenosis and thrombosis in haemodialysis fistulae and grafts: the surgeon's point of view [J].
Mickley, V .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2004, 19 (02) :309-311
[9]
Effect of a chitosan-based hemostatic dressing on blood loss and survival in a model of severe venous hemorrhage and hepatic injury in swine [J].
Pusateri, AE ;
McCarthy, SJ ;
Gregory, KW ;
Harris, RA ;
Cardenas, L ;
McManus, AT ;
Goodwin, CW .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2003, 54 (01) :177-182
[10]
Saran R, 2004, CONTRIB NEPHROL, V142, P14