Optimal electrocardiographically triggered phase for reducing motion artifact at electron-beam CT in the coronary artery

被引:18
作者
He, S
Dai, RP
Chen, Y
Bai, H
机构
[1] Chinese Acad Med Sci, Peking Union Med Coll, Cardiovasc Inst, Dept Radiol, Beijing 100037, Peoples R China
[2] Chinese Acad Med Sci, Fuwai Hosp, Peking Union Med Coll, Beijing 100037, Peoples R China
关键词
computed tomography (CT); artifact; electron beam; coronary vessels; CT;
D O I
10.1016/S1076-6332(03)80743-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. The authors performed this study to (a) investigate coronary movement with electron-beam computed tomography (CT) and (b) find the optimal electrocardiographic (ECG) triggering phase for eliminating motion artifact. Materials and Methods. One hundred fifty-one patients without arrhythmia were examined with electron-beam CT. First, movie scans were obtained to create displacement and velocity graphs of coronary artery movement. Then, a volume scan with an exposure time of 100 msec was obtained with various ECG trigger settings. Results. Movement patterns of coronary arteries varied with heart rate. Optimal triggering phase was before atrial systole (near 71% of the R-R interval) when heart rate was slower than 68 beats per minute and at ventricular end systole when heart rate was fast. Rate of severe motion artifacts decreased from 43% to 0% when triggering was altered from 80% of the R-R interval to the individual optimal value. Experimental values of the optimal phase at different heart rates were derived, and severe motion artifact was only 3.0% with these values. Conclusion, ECG triggering set according to the heart rate enables a great reduction in motion artifacts at electron-beam CT with a 100-msec exposure time. The results may have implications for magnetic resonance imaging of the coronary arte rv.
引用
收藏
页码:48 / 56
页数:9
相关论文
共 23 条
[1]  
Achenbach S, 1996, HERZ, V21, P106
[2]   Curved multiplanar reconstructions for the evaluation of contrast-enhanced electron beam CT of the coronary arteries [J].
Achenbach, S ;
Moshage, W ;
Ropers, D ;
Bachmann, K .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 170 (04) :895-899
[3]   Detection of high-grade restenosis after PTCA using contrast-enhanced electron beam CT [J].
Achenbach, S ;
Moshage, W ;
Bachmann, K .
CIRCULATION, 1997, 96 (09) :2785-2788
[4]  
Achenbach S, 1999, RADIOLOGY, V213P, P351
[5]   QUANTIFICATION OF CORONARY-ARTERY CALCIUM USING ULTRAFAST COMPUTED-TOMOGRAPHY [J].
AGATSTON, AS ;
JANOWITZ, WR ;
HILDNER, FJ ;
ZUSMER, NR ;
VIAMONTE, M ;
DETRANO, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) :827-832
[6]   CORONARY-ARTERY CALCIFICATION DETECTED WITH ULTRAFAST CT AS AN INDICATION OF CORONARY-ARTERY DISEASE - WORK IN PROGRESS [J].
BREEN, JF ;
SHEEDY, PF ;
SCHWARTZ, RS ;
STANSON, AW ;
KAUFMANN, RB ;
MOLL, PP ;
RUMBERGER, JA .
RADIOLOGY, 1992, 185 (02) :435-439
[7]   Intravenous three-dimensional coronary angiography using contrast enhanced electron beam computed tomography [J].
Budoff, MJ ;
Oudiz, RJ ;
Zalace, CP ;
Bakhsheshi, H ;
Goldberg, SL ;
French, WJ ;
Rami, TG ;
Brundage, BH .
AMERICAN JOURNAL OF CARDIOLOGY, 1999, 83 (06) :840-845
[8]   Evaluation of electron beam CT coronary angiography in healthy subjects [J].
Chernoff, DM ;
Ritchie, CJ ;
Higgins, CB .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1997, 169 (01) :93-99
[9]   Coronary MR angiography [J].
Duerinckx, AJ .
RADIOLOGIC CLINICS OF NORTH AMERICA, 1999, 37 (02) :273-+
[10]   Noninvasive coronary artery imaging using CT and MR imaging [J].
Duerinckx, AJ ;
Lipton, MJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 170 (04) :900-902