Simultaneous in vitro measurement of patellofemoral kinematics and forces following Oxford medial unicompartmental knee replacement

被引:30
作者
Price, A. J.
Oppold, P. T.
Murray, D. W.
Zavatsky, A. B.
机构
[1] Nuffield Orthopaed Ctr, Nuffield Dept Orthopaed Surg, NHS Trust, Oxford OX3 7LD, England
[2] Univ Oxford, Dept Engn Sci, Oxford OX1 3PJ, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2006年 / 88B卷 / 12期
关键词
D O I
10.1302/0301-620X.88B12.18306
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The Oxford medial unicompartmental knee replacement was designed to reproduce normal mobility and forces in the knee, but its detailed effect on the patellofemoral joint has not been studied previously. We have examined the effect on patellofemoral mechanics of the knee by simultaneously measuring patellofemoral kinematics and forces in 11 cadaver knee specimens in a supine leg-extension rig. Comparison was made between the intact normal knee and sequential unicompartmental and total knee replacement. Following medial mobile-bearing unicompartmental replacement in 11 knees, patellofemoral kinematics and forces did not change significantly from those in the intact knee across any measured parameter. In contrast, following posterior cruciate ligament retaining total knee replacement in eight knees, there were significant changes in patellofemoral movement and forces. The Oxford device appears to produce near-normal patellofemoral mechanics, which may partly explain the low incidence of complications with the extensor mechanism associated with clinical use.
引用
收藏
页码:1591 / 1595
页数:5
相关论文
共 34 条
[1]
Biomechanics of patellofemoral joint prostheses [J].
Amis, AA ;
Senavongse, W ;
Darcy, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (436) :20-29
[2]
Banks S A, 1997, Am J Knee Surg, V10, P261
[3]
Mechanism of anterior impingement damage in total knee arthroplasty [J].
Banks, SA ;
Harman, MK ;
Hodge, WA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A :37-42
[4]
BERGER RA, 2004, CLIN ORTHOP RELAT R, V28, P92
[5]
BLAND M, 2000, INTRO MED STAT, P175
[6]
Patellofemoral forces after total knee arthroplasty: effect of extensor moment arm [J].
Browne, C ;
Hermida, JC ;
Bergula, A ;
Colwell, CW ;
D'Lima, DD .
KNEE, 2005, 12 (02) :81-88
[7]
Emerson Jr R. H., 2002, CLIN ORTHOP RELAT R, DOI [10.1097/00003086-200211000-00011, DOI 10.1097/00003086-200211000-00011]
[8]
MECHANICS OF KNEE AND PROSTHESIS DESIGN [J].
GOODFELLOW, J ;
OCONNOR, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03) :358-369
[9]
GOODFELLOW J, 1992, J BONE JOINT SURG BR, V74, P1
[10]
A JOINT COORDINATE SYSTEM FOR THE CLINICAL DESCRIPTION OF 3-DIMENSIONAL MOTIONS - APPLICATION TO THE KNEE [J].
GROOD, ES ;
SUNTAY, WJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1983, 105 (02) :136-144