Band artifacts due to bulk motion

被引:55
作者
Storey, P
Chen, Q
Li, W
Edelman, RR
Prasad, PV
机构
[1] Northwestern Univ, Sch Med, Chicago, IL USA
[2] Evanston NW Healthcare, Dept Radiol, Evanston, IL USA
关键词
MRI; motion artifacts; parallel imaging; phase-encoding order; myocardium;
D O I
10.1002/mrm.10314
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Band artifacts due to bulk motion were investigated in images acquired with fast gradient echo sequences. A simple analytical calculation shows that the width of the artifacts has a square-root dependence on the velocity of the imaged object, the time taken to acquire each line of k-space and the field of view in the phase-encoding direction. The theory furthermore predicts that the artifact width can be reduced using parallel imaging by a factor equal to the square root of the acceleration parameter. The analysis and results are presented for motion in the phase- and frequency-encoding directions and comparisons are made between sequential and centric ordering. The theory is validated in phantom experiments, in which bulk motion is simulated in a controlled and reproducible manner by rocking the scan table back and forth along the bore axis. Preliminary cardiac studies in healthy human volunteers show that dark bands may be observed in the endocardium in images acquired with nonsegmented fast gradient echo sequences. The fact that the position of the bands changes with the phase-encoding direction suggests that they may be artifacts due to motion of the heart walls during the image acquisition period.
引用
收藏
页码:1028 / 1036
页数:9
相关论文
共 15 条
[1]   ACQUISITION ORDER AND MOTIONAL ARTIFACT REDUCTION IN SPIN WARP IMAGES [J].
DIXON, WT ;
BRUMMER, ME ;
MALKO, JA .
MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (01) :74-83
[2]  
HAACKE E M, 1986, Magnetic Resonance Imaging, V4, P359, DOI 10.1016/0730-725X(86)91046-5
[3]   PSEUDO-GATING - ELIMINATION OF PERIODIC MOTION ARTIFACTS IN MAGNETIC-RESONANCE-IMAGING WITHOUT GATING [J].
HAACKE, EM ;
LENZ, GW ;
NELSON, AD .
MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (02) :162-174
[4]   Phase-encode reordering to minimize errors caused by motion [J].
Macgowan, CK ;
Wood, ML .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (03) :391-398
[5]   Optimization of view ordering for motion artifact suppression [J].
Nguyen, TD ;
Ding, GL ;
Watts, R ;
Wang, Y .
MAGNETIC RESONANCE IMAGING, 2001, 19 (07) :951-957
[6]  
Pruessmann KP, 1999, MAGNET RESON MED, V42, P952, DOI 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO
[7]  
2-S
[8]   Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arrays [J].
Sodickson, DK ;
Manning, WJ .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (04) :591-603
[9]   Breath-hold three-dimensional contrast-enhanced coronary MR angiography: Motion-matched k-space sampling for reducing cardiac motion effects [J].
Wang, Y ;
Winchester, PA ;
Yu, L ;
Watts, R ;
Ding, GL ;
Lee, HM ;
Bergman, GW .
RADIOLOGY, 2000, 215 (02) :600-607
[10]   MOTIONAL PHASE ARTIFACTS IN FOURIER-TRANSFORM MRI [J].
WEDEEN, VJ ;
WENDT, RE ;
JEROSCHHEROLD, M .
MAGNETIC RESONANCE IN MEDICINE, 1989, 11 (01) :114-120