Determination of chromium in human lung fibroblast cells using a large bore-direct injection high-efficiency nebulizer with inductively coupled plasma mass spectrometry

被引:29
作者
McLean, JA
Acon, BW
Montaser, A [1 ]
Singh, J
Pritchard, DE
Patierno, SR
机构
[1] George Washington Univ, Dept Chem, Washington, DC 20052 USA
[2] George Washington Univ, Med Ctr, Dept Pharmacol, Washington, DC 20037 USA
关键词
large bore-direct injection high efficiency nebulizer; flow injection analysis; inductively coupled plasma mass spectrometry; chromium; cells; human lung fibroblasts;
D O I
10.1366/0003702001950120
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A novel method for the determination of chromium in suspensions of human lung fibroblast cells is described by using a large bore-direct injection high efficiency nebulizer (LB-DIHEN) with microscale how injection analysis and inductively coupled plasma mass spectrometric detection, Chromium(VI)-treated cells were first counted and then suspended in a phosphate buffer saline solution, With the use of the method of standard additions, the relative concentration of Cr in similar to 100 HLF cells/peak was determined at m/z = 50. Because the cells tend to clump and can yield inhomogeneities in the total number analyzed, Mg was used as an internal standard to compensate for the total cell mass. The level of Cr in HLF cells grown in a medium of 100 mu M Na2CrO4 for two hours Is on the order of 180 fg Cr/cell after correction for the number of cells in each injection.
引用
收藏
页码:659 / 663
页数:5
相关论文
共 40 条
[1]   A large bore-direct injection high efficiency nebulizer for inductively coupled plasma spectrometry [J].
Acon, BW ;
McLean, JA ;
Montaser, A .
ANALYTICAL CHEMISTRY, 2000, 72 (08) :1885-1893
[2]  
Bandura DR, 1999, ATOM SPECTROSC, V20, P69
[3]   A dynamic reaction cell for inductively coupled plasma mass spectrometry (ICP-DRC-MS) - Part 1. The rf-field energy contribution in thermodynamics of ion-molecule reactions [J].
Baranov, VI ;
Tanner, SD .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (08) :1133-1142
[4]   Ultratrace and isotope analysis of long-lived radionuclides by inductively coupled plasma quadrupole mass spectrometry using a direct injection high efficiency nebulizer [J].
Becker, JS ;
Dietze, HJ ;
McLean, JA ;
Montaser, A .
ANALYTICAL CHEMISTRY, 1999, 71 (15) :3077-3084
[5]  
Bollinger DS, 1999, ATOM SPECTROSC, V20, P60
[6]  
DOUGLAS DJ, 1989, CAN J SPECTROSC, V34, P38
[7]  
Freshney RI., 1994, Culture of animal cells : A manual of basic technique, V3rd, DOI DOI 10.1002/9780470649367
[8]  
GOYER RA, 1995, HDB EXPT PHARM, V115
[9]  
Hanada T, 1998, CANCER, V83, P930, DOI 10.1002/(SICI)1097-0142(19980901)83:5<930::AID-CNCR19>3.0.CO
[10]  
2-W