The effect of analyte concentration range on measurement errors obtained by NIR spectroscopy

被引:25
作者
Riley, MR [1 ]
Crider, HM [1 ]
机构
[1] Univ Arizona, Dept Agr & Biosyst Engn, Tucson, AZ 85721 USA
关键词
NIR spectroscopy; bioreactor monitoring; animal cell culture;
D O I
10.1016/S0039-9140(00)00403-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Near infrared spectroscopy (NIRS) was employed to quantify five compounds, ammonium, glucose, glutamate, glutamine, and lactate, in conditions similar to those obtained in animal cell cultivations over varying ranges of analyte concentrations. These components represent the primary nutrients and wastes of animal cells for which such noninvasive monitoring schemes are required for development of accurate control schemes. Ideal cultivation conditions involve maintaining concentrations of these components as low as 1 mM each, however, it is not known if measurements of these compounds can be accurately accomplished at such a low level. We have found that NIPS measurements of these analytes over narrow and low (0-1 mM) concentration ranges yield measurement errors of roughly 11% of the concentration range. By contrast, wide concentration ranges (0-30 mM) yield measurement errors of roughly 1.6% of the concentration range. Decreasing the concentration range over which an analyte is quantified in four out of five cases decreases the optimal spectral range by 100 cm (-1) for measurement by partial least squares regression analysis. There appears a similarity in the ratio of (standard error of prediction (SEP)/concentration range) which may provide an estimation of the anticipated SEP to be obtained for measurement over a new concentration range. It was found that for the five analytes evaluated here, the ratio of SEP to concentration range divided by that obtained for a second concentration range is equal to a fairly constant value of 6.6. This relationship was found to be followed reasonably well by an extensive number of measurement results reported in the literature for similar conditions. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:473 / 484
页数:12
相关论文
共 34 条
[1]  
ASTM, 1995, STAND PRACT INFR MUL
[2]   NONINVASIVE METHOD FOR MONITORING ETHANOL IN FERMENTATION PROCESSES USING FIBEROPTIC NEAR-INFRARED SPECTROSCOPY [J].
CAVINATO, AG ;
MAYES, DM ;
GE, ZH ;
CALLIS, JB .
ANALYTICAL CHEMISTRY, 1990, 62 (18) :1977-1982
[3]   Simultaneous measurements of glucose, glutamine, ammonia, lactate, and glutamate in aqueous solutions by near-infrared spectroscopy [J].
Chung, H ;
Arnold, MA ;
Rhiel, M ;
Murhammer, DW .
APPLIED SPECTROSCOPY, 1996, 50 (02) :270-276
[4]   SIMULTANEOUS MEASUREMENT OF GLUCOSE AND GLUTAMINE IN AQUEOUS-SOLUTIONS BY NEAR-INFRARED SPECTROSCOPY [J].
CHUNG, HI ;
ARNOLD, MA ;
RHIEL, M ;
MURHAMMER, DW .
APPLIED BIOCHEMISTRY AND BIOTECHNOLOGY, 1995, 50 (02) :109-125
[5]   AN ENERGETICALLY STRUCTURED MODEL OF MAMMALIAN-CELL METABOLISM .1. MODEL DEVELOPMENT AND APPLICATION TO STEADY-STATE HYBRIDOMA CELL-GROWTH IN CONTINUOUS-CULTURE [J].
DIMASI, D ;
SWARTZ, RW .
BIOTECHNOLOGY PROGRESS, 1995, 11 (06) :664-676
[6]   In situ monitoring of an Escherichia coli fermentation using a diamond composition ATR probe and mid-infrared spectroscopy [J].
Doak, DL ;
Phillips, JA .
BIOTECHNOLOGY PROGRESS, 1999, 15 (03) :529-539
[7]   MATHEMATICAL DESCRIPTIONS OF HYBRIDOMA CULTURE KINETICS .3. SIMULATION OF FED-BATCH BIOREACTORS [J].
GLACKEN, MW ;
HUANG, C ;
SINSKEY, AJ .
JOURNAL OF BIOTECHNOLOGY, 1989, 10 (01) :39-65
[8]   PARTIAL LEAST-SQUARES METHODS FOR SPECTRAL ANALYSES .1. RELATION TO OTHER QUANTITATIVE CALIBRATION METHODS AND THE EXTRACTION OF QUALITATIVE INFORMATION [J].
HAALAND, DM ;
THOMAS, EV .
ANALYTICAL CHEMISTRY, 1988, 60 (11) :1193-1202
[9]   The use of NIR spectroscopy in monitoring and controlling bioprocesses [J].
Hagman, A ;
Sivertsson, P .
PROCESS CONTROL AND QUALITY, 1998, 11 (02) :125-128
[10]   Near-infrared spectroscopic determination of acetate, ammonium, biomass, and glycerol in an industrial Escherichia coli fermentation [J].
Hall, JW ;
McNeil, B ;
Rollins, MJ ;
Draper, I ;
Thompson, BG ;
Macaloney, G .
APPLIED SPECTROSCOPY, 1996, 50 (01) :102-108