Segmental bioelectrical impedance analysis to assess perioperative fluid changes

被引:27
作者
Bracco, D
Berger, MM
Revelly, JP
Schütz, Y
Frascarolo, P
Chioléro, R
机构
[1] Univ Lausanne Hosp, Surg Intens Care Unit, Lausanne, Switzerland
[2] Univ Lausanne Hosp, Dept Anesthesiol, Lausanne, Switzerland
[3] Univ Lausanne, Fac Med, Inst Physiol, CH-1005 Lausanne, Switzerland
关键词
bioelectrical impedance analysis; body composition; fluid resuscitation; fluid balance; critically ill; surgery; edema;
D O I
10.1097/00003246-200007000-00034
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: Perioperative fluid accumulation determination is a challenge for the clinician, Bioelectrical impedance analysis (BIA) is a noninvasive method based on the electrical properties of tissues, which can assess body fluid compartments, The study aimed at assessing their changes in three types of surgery (thoracic, abdominal, and intracranial) requiring various regimens of fluid administration, Design: Prospective descriptive trial, Patients: A total of 26 patients scheduled for elective surgery were separated into three groups according to site of surgery: thoracic (n = 8), abdominal aortic (n = 8), and brain surgery (n = 10), Setting: University teaching hospital, Intervention: None. Measurements: Whole body, segmental (arm, trunk, and legs) BIA at multiple frequency (0.5, 50, 100 kHz) was used to assess perioperative fluid accumulation after surgery, The fluid balances were calculated from the charts, Results: The patients were aged 62 +/- 4 yrs, Fluid balances were 4.8 +/- 1.0 L, 4.1 +/- 0.5 L, and 1.9 +/- 0.3 L, respectively, in the three groups, In trunk surgery patients, fluid accumulation was detected as a drop in impedance in the operated area at all frequencies, In the operated area, there was an expansion of both intra- and extracellular compartments, A reduction in high frequencies' impedance in the legs was only detected after aortic surgery, Fluid accumulation and trunk impedance changes were strongly correlated, Neurosurgery only induced minor body fluid changes, Conclusions: Segmental BIA is able to detect and localize perioperative fluid accumulation, It may become a bedside tool to quantify and to localize fluid accumulation.
引用
收藏
页码:2390 / 2396
页数:7
相关论文
共 32 条
[1]   BIOELECTRIC IMPEDANCE PHASE-ANGLE AND BODY-COMPOSITION [J].
BAUMGARTNER, RN ;
CHUMLEA, WC ;
ROCHE, AF .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1988, 48 (01) :16-23
[2]  
BESHYAH SA, 1995, AM J CLIN NUTR, V61, P1186
[3]   Segmental body composition assessed by bioelectrical impedance analysis and DEXA in humans [J].
Bracco, D ;
Thiebaud, D ;
Chiolero, RL ;
Landry, M ;
Burckhardt, P ;
Schutz, Y .
JOURNAL OF APPLIED PHYSIOLOGY, 1996, 81 (06) :2580-2587
[4]   Bedside determination of fluid accumulation after cardiac surgery using segmental bioelectrical impedance [J].
Bracco, D ;
Revelly, JP ;
Berger, MM ;
Chioléro, RL .
CRITICAL CARE MEDICINE, 1998, 26 (06) :1065-1070
[5]  
BRUMMER RJ, 1992, EUR J CLIN NUTR, V46, P117
[6]   MULTIFREQUENCY BIOELECTRICAL-IMPEDANCE ESTIMATES THE DISTRIBUTION OF BODY-WATER [J].
CHA, KC ;
CHERTOW, GM ;
GONZALEZ, J ;
LAZARUS, JM ;
WILMORE, DW .
JOURNAL OF APPLIED PHYSIOLOGY, 1995, 79 (04) :1316-1319
[7]   ASSESSMENT OF CHANGES IN BODY-WATER BY BIOIMPEDANCE IN ACUTELY ILL SURGICAL PATIENTS [J].
CHIOLERO, RL ;
GAY, LJ ;
COTTING, J ;
GURTNER, C ;
SCHUTZ, Y .
INTENSIVE CARE MEDICINE, 1992, 18 (06) :322-326
[8]  
CHUMLEA WC, 1990, ADV IN VIVO BODY COM, P375
[9]   IMPROVED PREDICTION OF EXTRACELLULAR AND TOTAL-BODY WATER USING IMPEDANCE LOCI GENERATED BY MULTIPLE FREQUENCY BIOELECTRICAL IMPEDANCE ANALYSIS [J].
CORNISH, BH ;
THOMAS, BJ ;
WARD, LC .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (03) :337-346
[10]   MULTIFREQUENCY IMPEDANCE FOR THE PREDICTION OF EXTRACELLULAR WATER AND TOTAL-BODY WATER [J].
DEURENBERG, P ;
TAGLIABUE, A ;
SCHOUTEN, FJM .
BRITISH JOURNAL OF NUTRITION, 1995, 73 (03) :349-358