Understanding wear in dual mobility total hip replacement: first generation explant wear patterns

被引:25
作者
Boyer, Bertrand [1 ,2 ]
Neri, Thomas [2 ]
Geringer, Jean [1 ]
Di Iorio, Alexandre [2 ]
Philippot, Remi [2 ]
Farizon, Frederic [2 ]
机构
[1] Univ Lyon St Etienne, INSERM, U1059, Mines St Etienne,Unite SAINBIOSE, Lyon, France
[2] Univ Jean Monnet, Fac Med J Lisfranc, Lab Physiol Exercice, F-42055 St Etienne 2, France
关键词
Dual mobility; Wear; Implant retrieval; THR; Wear patterns; INTRA-PROSTHETIC DISLOCATION; FOLLOW-UP; ARTHROPLASTY; TERM; SOCKET; CUP;
D O I
10.1007/s00264-016-3362-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Purpose Dual mobility was introduced by Pr.G. Bousquet to improve the dislocation risk of total hip arthroplasty (THR). The wear mechanisms of this implant remain to be understood. Could enhanced explant wear analysis explain liner wear behaviour? Methods Sixty-six explanted liners were selected. Non-destructive 3D scanning and heat colour treatment allowed reporting wear patterns on both convex and concave sides of the liners. Results For many liners, liner convexity wear was found to be homogeneous, and linked to a longer implantation time. A few liners had signs of blocking of the convexity, leading to excessive internal damage. Intra prosthetic dislocations were found to only have internal circular wear without liner penetration. Conclusion Heat colour treatment showed its effectiveness for highlighting wear patterns. Optimal convexity mobility seemed to be leading to a longer implantation time, with decreased contact stresses on concavity. The retentive rimshould have a distinct wear analysis.
引用
收藏
页码:529 / 533
页数:5
相关论文
共 22 条
[1]
Wear paths produced by individual hip-replacement patients - A large-scale, long-term follow-up study [J].
Bennett, D. ;
Humphreys, L. ;
O'Brien, S. ;
Kelly, C. ;
Orr, J. F. ;
Beverland, D. E. .
JOURNAL OF BIOMECHANICS, 2008, 41 (11) :2474-2482
[2]
Primary total hip arthroplasty with dual mobility socket to prevent dislocation: a 22-year follow-up of 240 hips [J].
Boyer, Bertrand ;
Philippot, Remi ;
Geringer, Jean ;
Farizon, Frederic .
INTERNATIONAL ORTHOPAEDICS, 2012, 36 (03) :511-518
[3]
A comparative and retrospective study of three hundred and twenty primary Charnley type hip replacements with a minimum follow up of ten years to assess wether a dual mobility cup has a decreased dislocation risk [J].
Caton, Jacques H. ;
Prudhon, Jean Louis ;
Ferreira, Andr ;
Aslanian, Thierry ;
Verdier, Regis .
INTERNATIONAL ORTHOPAEDICS, 2014, 38 (06) :1125-1129
[4]
OPTIMUM SIZE OF PROSTHETIC HEADS IN RELATION TO WEAR OF PLASTIC SOCKETS IN TOTAL REPLACEMENT OF HIP [J].
CHARNLEY, J ;
KAMANGAR, A ;
LONGFIELD, MD .
MEDICAL & BIOLOGICAL ENGINEERING, 1969, 7 (01) :31-+
[5]
A literature review of the association between wear rate and osteolysis in total hip arthroplasty [J].
Dumbleton, JH ;
Manley, MT ;
Edidin, AA .
JOURNAL OF ARTHROPLASTY, 2002, 17 (05) :649-661
[6]
Long-term results of an uncemented straight femoral shaft prosthesis [J].
Eingartner, C ;
Heigele, T ;
Volkmann, R ;
Weise, K .
HIP INTERNATIONAL, 2006, 16 (01) :23-32
[7]
Engh CA, 2001, CLIN ORTHOP RELAT R, P137
[8]
Uncemented total hip arthroplasty for primary osteoarthritis in young patients - A mid- to long-term follow-up study from the Finnish Arthroplasty Register [J].
Eskelinen, A ;
Remes, V ;
Helenius, I ;
Pulkkinen, P ;
Nevalainen, J ;
Paavolainen, P .
ACTA ORTHOPAEDICA, 2006, 77 (01) :57-70
[9]
Intra-prosthetic dislocation of dual-mobility cups after total hip arthroplasty: potential causes from a clinical and biomechanical perspective [J].
Fabry, Christian ;
Langlois, Jean ;
Hamadouche, Moussa ;
Bader, Rainer .
INTERNATIONAL ORTHOPAEDICS, 2016, 40 (05) :901-906
[10]
Understanding the dual mobility concept for total hip arthroplasty. Investigations on a multiscale analysis-highlighting the role of arthrofibrosis [J].
Geringer, J. ;
Boyer, B. ;
Farizon, F. .
WEAR, 2011, 271 (9-10) :2379-2385