Validation of left ventricular end-diastolic volume from stroke volume and ejection fraction

被引:8
作者
Hart, JP [1 ]
Cabreriza, SE [1 ]
Gallup, CG [1 ]
Hsu, D [1 ]
Spotnitz, HM [1 ]
机构
[1] Columbia Univ Coll Phys & Surg, Dept Surg, New York, NY 10032 USA
关键词
D O I
10.1097/00002480-200211000-00014
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The present study examines an innovative approach to measurement of left ventricular (LV) end-diastolic volume (LVEDV). Measurement of LVEDV is fundamental to the assessment of intraoperative systolic and diastolic LV function. We compared steady state LVEDV values obtained from stroke volume (SV) and ejection fraction (EF) with echocardiographic and postmortem LVEDV measurements. Five anesthetized pigs (40-45 kg) underwent median sternotomy and pericardiotomy. A transit time ultrasonic flow probe was placed on the ascending aorta to provide cardiac output. A micromanometer provided LV end-diastolic pressure. LV short axis cross sectional echocardiograms and electrocardiograms were also obtained. LV end-diastolic area (LVEDA) and end-systolic area (LVESA) were measured to obtain EF. LVEDV-end was calculated from cardiac output, heart rate, and EF. LVEDVecho was determined using a three-plane echocardiography model. Postmortem (LVEDVpmlvv) volumes were also measured. LVEDVsv/ef correlated well with volumes obtained by echocardiography (r(2) = 0.92) and postmortem (r(2) = 0.73) measurements. Values of p < 0.05 indicated significant linearity of LVEDA-LVEDVsv/ef (r(2) =0.93), LVEDA-LVEDVecho (r(2) = 0.96), and LVEDA-LVEDVpmlvv (r(2) = 0.81) relationships. Determination of LVEDV from SV and EF is valid and may facilitate real-time determination of LV mechanics
引用
收藏
页码:654 / 657
页数:4
相关论文
共 24 条
[1]
Validation of right and left ventricular conductance and echocardiography for cardiac function studies [J].
Amirhamzeh, MMR ;
Dean, DA ;
Jia, CX ;
Cabreriza, SE ;
Yano, OJ ;
Burkhoff, D ;
Spotnitz, HM .
ANNALS OF THORACIC SURGERY, 1996, 62 (04) :1104-1109
[2]
CONTINUOUS MEASUREMENT OF LEFT-VENTRICULAR VOLUME IN ANIMALS AND HUMANS BY CONDUCTANCE CATHETER [J].
BAAN, J ;
VANDERVELDE, ET ;
DEBRUIN, HG ;
SMEENK, GJ ;
KOOPS, J ;
VANDIJK, AD ;
TEMMERMAN, D ;
SENDEN, J ;
BUIS, B .
CIRCULATION, 1984, 70 (05) :812-823
[3]
FACTORS CONFOUNDING IMPEDANCE CATHETER VOLUME MEASUREMENTS INVITRO [J].
BIELEFELD, MR ;
CABRERIZA, SE ;
SPOTNITZ, HM .
ANNALS OF THORACIC SURGERY, 1993, 55 (06) :1534-1539
[4]
CABRERIZA SE, 1995, ASAIO J, V41, pM669, DOI 10.1097/00002480-199507000-00095
[5]
CANINE LEFT-VENTRICULAR VOLUME AND MASS DURING THORACOTOMY BY TWO-DIMENSIONAL ECHOCARDIOGRAPHY - INCREASED VENTRICULAR MASS AFTER ISCHEMIA AND REPERFUSION [J].
COLLINS, RH ;
HAASLER, GB ;
KRUG, JH ;
COLT, C ;
SPOTNITZ, AJ ;
MARTIN, EC ;
SPOTNITZ, HM .
JOURNAL OF SURGICAL RESEARCH, 1982, 33 (04) :294-304
[6]
Validation study of a new transit time ultrasonic flow probe for continuous great vessel measurements [J].
Dean, DA ;
Jia, CX ;
Cabreriza, SE ;
DAlessandro, DA ;
Dickstein, ML ;
Sardo, MJ ;
Chalik, N ;
Spotnitz, HM .
ASAIO JOURNAL, 1996, 42 (05) :M671-M676
[7]
LEFT-VENTRICULAR EJECTION FRACTION DURING CARDIAC-SURGERY - A TWO-DIMENSIONAL ECHOCARDIOGRAPHIC STUDY [J].
DUBROFF, JM ;
CLARK, MB ;
WONG, CYH ;
SPOTNITZ, AJ ;
COLLINS, RH ;
SPOTNITZ, HM .
CIRCULATION, 1983, 68 (01) :95-103
[8]
HAASLER GB, 1985, J THORAC CARDIOV SUR, V90, P430
[9]
Conductance artifacts in a novel in vitro model of ventriculothoracic electrical coupling [J].
Hart, JP ;
Cabreriza, SE ;
Dean, DA ;
Jia, CX ;
Spotnitz, HM .
ASAIO JOURNAL, 1999, 45 (01) :64-68
[10]
RELATION BETWEEN LONGITUDINAL, CIRCUMFERENTIAL, AND OBLIQUE SHORTENING AND TORSIONAL DEFORMATION IN THE LEFT-VENTRICLE OF THE TRANSPLANTED HUMAN-HEART [J].
INGELS, NB ;
HANSEN, DE ;
DAUGHTERS, GT ;
STINSON, EB ;
ALDERMAN, EL ;
MILLER, DC .
CIRCULATION RESEARCH, 1989, 64 (05) :915-927