An SPR immunosensor for human cardiac troponin T using specific binding avidin to biotin at carboxymethyldextran-modified gold chip

被引:91
作者
Dutra, Rosa Fireman
Kubota, Lauro Tatsuo
机构
[1] Univ Estadual Campinas, Inst Quim, Dept Quim Analit, BR-13083970 Campinas, SP, Brazil
[2] Univ Fed Pernambuco, Inst Ciencias Biol, Dept Pathol, BR-50100130 Recife, PE, Brazil
关键词
troponin T; SPR; biotin-avidin; myocardial infarction; biosensor;
D O I
10.1016/j.cca.2006.07.029
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The cardiac troponin T (TnT) is a cardiospecific, highly sensitive marker for myocardial damage and is immediately released to bloodstream during the acute myocardial infarction (AMI). Surface plasmon resonance (SPR) immunosensor was developed for the quick detection of human cTnT in real time. Methods: An SPR sensor streptavidin terminated self-assembled monolayer, which was used to binding biotinylated anti troponin T monoclonal antibodies, was developed. Results: The cTnT was determined from successive injections with a linear range from 0.03 up to 6.5 ng/ml. The detection limit was 0.01 ng/ml corresponding to a resonant angle change of 1.28 millidegrees. The system presented a good repeatability with 3.4% of variation between run after regeneration of the coated surface with a solution of 1% SDS. The effects of blocking of non-specific adsorption using different solutions were studied. It was possible to measure the cTnT without dilution of the human serum with good specificity and reproducibility. Conclusion: This sensor is practical and offer quick response in interval of 800 s. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:114 / 120
页数:7
相关论文
共 23 条
[1]  
Apple FS, 2001, CLIN CHEM, V47, P377
[2]   Real-Time, label-free monitorine, of tumor antigen and serum antibody interactions [J].
Campagnolo, C ;
Meyers, KJ ;
Ryan, T ;
Atkinson, RC ;
Chen, YT ;
Scanlan, MJ ;
Ritter, G ;
Old, LJ ;
Batt, CA .
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 2004, 61 (03) :283-298
[3]   Application of SPR biosensor for medical diagnostics of human hepatitis B virus (hHBV) [J].
Chung, JW ;
Kim, SD ;
Bernhardt, R ;
Pyun, JC .
SENSORS AND ACTUATORS B-CHEMICAL, 2005, 111 :416-422
[4]   Determination of association rate constants by an optical biosensor using initial rate analysis [J].
Edwards, PR ;
Leatherbarrow, RJ .
ANALYTICAL BIOCHEMISTRY, 1997, 246 (01) :1-6
[5]   Highly sensitive regenerable immunosensor for label-free detection of 2,4-dichlorophenoxyacetic acid at ppb levels by using surface plasmon resonance imaging [J].
Gobi, KV ;
Tanaka, H ;
Shoyama, Y ;
Miura, N .
SENSORS AND ACTUATORS B-CHEMICAL, 2005, 111 :562-571
[6]   Clinical application of rapid quantitative determination of cardiac troponin-T in an emergency department setting [J].
Herkner, H ;
Waldenhofer, U ;
Laggner, AN ;
Müllner, M ;
Oschatz, E ;
Spitzauer, S ;
Gamper, G ;
Bur, A ;
Hirschl, MM .
RESUSCITATION, 2001, 49 (03) :259-264
[7]   Detection of electrochemical enzymatic reactions by surface plasmon resonance measurement [J].
Iwasaki, Y ;
Horiuchi, T ;
Niwa, O .
ANALYTICAL CHEMISTRY, 2001, 73 (07) :1595-1598
[8]  
Klein G, 1998, WIEN KLIN WOCHENSCHR, V110, P40
[9]   Formation of a self-assembled phenylboronic acid monolayer and its application toward developing a surface plasmon resonance-based monosaccharide sensor [J].
Lee, M ;
Kim, TI ;
Kim, KH ;
Kim, JH ;
Choi, MS ;
Choi, HJ ;
Koh, K .
ANALYTICAL BIOCHEMISTRY, 2002, 310 (02) :163-170
[10]   Fabrication of self-assembled protein A monolayer and its application as an immunosensor [J].
Lee, W ;
Oh, BK ;
Bae, YM ;
Paek, SH ;
Lee, WH ;
Choi, JW .
BIOSENSORS & BIOELECTRONICS, 2003, 19 (03) :185-192