Microchip capillary electrophoresis for point-of-care analysis of lithium

被引:50
作者
Vrouwe, Elwin X.
Luttge, Regina
Vermes, Istvan
van den Berg, Albert
机构
[1] Univ Twente, MESA, NL-7500 AE Enschede, Netherlands
[2] Univ Twente, Inst Nanotechnol, NL-7500 AE Enschede, Netherlands
[3] Hosp Grp, Med Spectrum Twente, Dept Clin Chem, Enschede, Netherlands
关键词
D O I
10.1373/clinchem.2007.073726
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Microchip capillary electrophoresis (CE) is a promising method for chemical analysis of complex samples such as whole blood. We evaluated the method for point-of-care testing of lithium. Methods: Chemical separation was performed on standard glass microchip CE devices with a conductivity detector as described in previous work. Here we demonstrate a new sample-to-chip interface. Initially, we took a glass capillary as a sample collector for whole blood from a finger stick. In addition, we designed a novel disposable sample collector and tested it against the clinical standard at the hospital (Medisch Spectrum Twente). Both types of collectors require < 10 mu L of test fluid. The collectors contain an integrated filter membrane, which prevents the transfer of blood cells into the microchip. The combination of such a sample collector with microchip CE allows point-of-care measurements without the need for off-chip sample treatment. This new on-chip protocol was verified against routine lithium testing of 5 patients in the hospital. Results: Sodium, lithium, magnesium, and calcium were separated in < 20 s. The detection limit for lithium was 0.15 mmol/L. Conclusions: The new microchip CE system provides a convenient and rapid method for point-of-care testing of electrolytes in serum and whole blood. (c) 2007 American Association for Clinical Chemistry
引用
收藏
页码:117 / 123
页数:7
相关论文
共 29 条
[1]  
AMDISEN A, 1975, DAN MED BULL, V22, P277
[2]   Inorganic pharmacology of lithium [J].
Birch, NJ .
CHEMICAL REVIEWS, 1999, 99 (09) :2659-2682
[3]   Determination of amphetamines in human whole blood by capillary electrophoresis with photodiode array detection [J].
Boatto, G ;
Faedda, MV ;
Pau, A ;
Asproni, B ;
Menconi, S ;
Cerri, R .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2002, 29 (06) :1073-1080
[4]  
Cheng J, 1998, TOP CURR CHEM, V194, P215
[5]   FIELD AMPLIFIED SAMPLE INJECTION IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS [J].
CHIEN, RL ;
BURGI, DS .
JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2) :141-152
[6]   Microchip-based capillary electrophoresis of human serum proteins [J].
Colyer, CL ;
Mangru, SD ;
Harrison, DJ .
JOURNAL OF CHROMATOGRAPHY A, 1997, 781 (1-2) :271-276
[7]   Fluidics and sample handling in clinical chemical analysis [J].
Cunningham, DD .
ANALYTICA CHIMICA ACTA, 2001, 429 (01) :1-18
[8]   Capillary electrophoresis of proteins 2003-2005 [J].
Dolnik, V .
ELECTROPHORESIS, 2006, 27 (01) :126-141
[9]  
ERICKSON KA, 1993, CLIN CHEM, V39, P283
[10]   Direct detection of renal function markers using microchip CE with pulsed electrochemical detection [J].
Garcia, CD ;
Henry, CS .