Simple and rapid determination of amphetamine, methamphetamine, and their methylenedioxy derivatives in urine by automated in-tube solid-phase microextraction coupled with liquid chromatography-electrospray ionization mass spectrometry

被引:83
作者
Kataoka, H [1 ]
Lord, HL [1 ]
Pawliszyn, J [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
关键词
D O I
10.1093/jat/24.4.257
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and rapid method for the determination of amphetamine, methamphetamine, and their 3,4-methylenedioxy derivatives in urine samples was developed using automated in-tube solid-phase microextraction (SPME) coupled with liquid chromatography-electrospray ionization mass spectrometry (LC-ESI-MS). In-tube SPME is an extraction technique for organic compounds in aqueous samples in which analytes are extracted from the sample directly into an open tubular capillary by repeated draw/eject cycles of sample solution. LC-MS analyses of stimulants were initially performed by liquid injection onto an LC column to determine spectra. Five stimulants tested in this study gave very simple ESI mass spectra, and strong signals corresponding to [M+H]+ were observed for all stimulants. The stimulants were well separated with a Supelcosil LC-CN column using acetonitrile/50mM ammonium acetate (15:85) as a mobile phase. In order to optimize the extraction of stimulants, several in-tube SPME parameters were examined. The optimum extraction conditions were 15 draw/eject cycles of 35 μL of sample in 50mM Tris-HCl (pH 8.5) at a flow rate of 100 μL/min using an Omegawax 250 capillary column. The stimulants extracted by the capillary were easily desorbed by mobile phase flow, and carryover of stimulants was not observed. Using in-tube SPME- LC-ESI-MS with selected ion monitoring, the calibration curves of stimulants were linear in the range from 2 to 100 ng/mL with correlation coefficients above 0.9985 (n = 18) and detection limits (S/N = 3) of 0.38-0.82 ng/mL. This method was successfully applied to the analysis of human urine samples without interference peaks. The recoveries of stimulants spiked into urine samples were above 81%.
引用
收藏
页码:257 / 265
页数:9
相关论文
共 60 条
[1]  
[Anonymous], JPN J FORENSIC TOXIC
[2]   Determination of a non-ionic surfactant by solid-phase microextraction coupled with high-performance liquid chromatography and on-line derivatization [J].
Aranda, R ;
Burk, RC .
JOURNAL OF CHROMATOGRAPHY A, 1998, 829 (1-2) :401-406
[3]   SOLID-PHASE MICROEXTRACTION WITH THERMAL-DESORPTION USING FUSED-SILICA OPTICAL FIBERS [J].
ARTHUR, CL ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1990, 62 (19) :2145-2148
[4]  
Battu C, 1998, J CHROMATOGR SCI, V36, P1
[5]   Determination of phenylisothiocyanate derivatives of amphetamine and its analogues in biological fluids by HPLC-APCI-MS or DAD [J].
Bogusz, MJ ;
Kala, M ;
Maier, RD .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1997, 21 (01) :59-69
[6]   Solid-phase microextraction coupled with high-performance liquid chromatography for the determination of alkylphenol ethoxylate surfactants in water [J].
BoydBoland, AA ;
Pawliszyn, JB .
ANALYTICAL CHEMISTRY, 1996, 68 (09) :1521-1529
[7]  
BUCHANAN JF, 1988, MED TOXICOL ADV DRUG, V3, P1
[8]   Quantitative and qualitative analysis of MDMA, MDEA, MA and amphetamine in urine by head-space/solid phase micro-extraction (SPME) and GC/MS [J].
Centini, F ;
Masti, A ;
Comparini, IB .
FORENSIC SCIENCE INTERNATIONAL, 1996, 83 (03) :161-166
[9]   SOLID-PHASE MICROEXTRACTION COUPLED TO HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
CHEN, J ;
PAWLISZYN, JB .
ANALYTICAL CHEMISTRY, 1995, 67 (15) :2530-2533
[10]   Comparison of new solid-phase extraction methods for chromatographic identification of drugs in clinical toxicological analysis [J].
Degel, F .