Evaluation of metal allergies in patients with coronary stents

被引:66
作者
Hillen, U
Haude, M
Erbel, R
Goos, M
机构
[1] Univ Essen Gesamthsch, Dept Dermatol, D-45122 Essen, Germany
[2] Univ Essen Gesamthsch, Dept Cardiol, D-45122 Essen, Germany
关键词
316 L stainless steel; coronary stent; delayed hypersensitivity reaction; in-stent restenosis; metal allergy; nickel; patch test; stent;
D O I
10.1034/j.1600-0536.2002.470607.x
中图分类号
R392 [医学免疫学];
学科分类号
100102 ;
摘要
In-stent restenosis is a complication which impairs the success of coronary stenting. Recently, it was supposed that a delayed hypersensitivity reaction to nickel and molybdenum might be one of the triggering factors in in-stent restenosis. We have analyzed the data collected in our centre with respect to this hypothesis. Altogether, 34 patients were investigated (24 male, 10 female). Patch tests were performed with the standard series of the German Contact Dermatitis Research Group and a metal series containing the metal components of 316 L stainless steel. A positive patch test reaction to nickel was observed in 4 (11.8%) patients. None of the patients showed sensitization to the other metals. Retrospective analysis was performed in 20 patients: 2 of these patients had a positive patch test reaction to nickel, one of whom had an in-stent restenosis, and the other not. Restenoses were predominantly observed in patients with negative patch test reactions to nickel (6/18 patients). Out of the patients who were investigated prospectively only one showed sensitization to nickel. Restenosis was observed in 2 patients: neither patient had nickel allergy. Although it still cannot be excluded that metal allergy may play a role in the restenosis process in coronary stenting, there is at present little evidence for it.
引用
收藏
页码:353 / 356
页数:4
相关论文
共 22 条
  • [1] Brasch J, 1998, HAUTARZT, V49, P184, DOI 10.1007/s001050050725
  • [2] BREHLER R, 1995, DTSCH DERMATOLOGE, V43, P688
  • [3] CARLSSON A, 1989, ACTA DERM-VENEREOL, V69, P62
  • [4] DEGROOT AC, 2001, TXB CONTACT DERMATIT, P1037
  • [5] Pathology of acute and chronic coronary stenting in humans
    Farb, A
    Sangiorgi, G
    Carter, AJ
    Walley, VM
    Edwards, WD
    Schwartz, RS
    Virmani, R
    [J]. CIRCULATION, 1999, 99 (01) : 44 - 52
  • [6] Morphological predictors of restenosis after coronary stenting in humans
    Farb, A
    Weber, DK
    Kolodgie, FD
    Burke, AP
    Virmani, R
    [J]. CIRCULATION, 2002, 105 (25) : 2974 - 2980
  • [7] NICKEL SENSITIVITY AND THE IMPLANTATION OF ORTHOPEDIC PROSTHESES
    GAWKRODGER, DJ
    [J]. CONTACT DERMATITIS, 1993, 28 (05) : 257 - 259
  • [8] Metal-induced generalized pruriginous dermatitis and endovascular surgery
    Giménez-Arnau, A
    Riambau, V
    Serra-Baldrich, E
    Camarasa, JG
    [J]. CONTACT DERMATITIS, 2000, 43 (01) : 35 - 40
  • [9] Greim H., 2000, GESUNDHEITSSCHADLICH, V31
  • [10] Metal sensitivity in patients with orthopaedic implants
    Hallab, N
    Merritt, K
    Jacobs, JJ
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (03) : 428 - 436