Improvement on simultaneous determination of chromium species in aqueous solution by ion chromatography and chemiluminescence detection

被引:67
作者
Gammelgaard, B [1 ]
Liao, YP [1 ]
Jons, O [1 ]
机构
[1] BEIJING UNIV,BEIJING 100871,PEOPLES R CHINA
关键词
chromium speciation; ion chromatography; flow injection analysis; chemiluminescence;
D O I
10.1016/S0003-2670(97)00421-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sensitive method for the simultaneous determination of chromium(III) and chromium(VI) was chromatography and chemiluminescence detection. Two Dionex ion-exchange guard columns in series, CG5 and AG7, were used to separate chromium(III) from chromium(VI). Chromium(VI) was reduced by potassium sulphite, whereupon both species were detected by use of the luminol-hydrogen peroxide chemiluminescence system. Parameters affecting retention times and resolution of the separator columns, such as eluent pH, eluent composition, reductant pH and concentration, and flow rates were optimized. Furthermore, the stabilities of reductant and luminol solutions were studied. The linear range of the calibration curve for chromium(III) and chromium(VI) was 1-400 mu g l(-1). The detection limit was 0.12 mu g l(-1) for chromium(III) and 0.09 mu g l(-1) for chromium(VI), respectively. The precision at the 20 mu g l(-1) level was 1.4% for chromium(III) and 2.5% for chromium(VI), respectively. The accuracy of the chromium(III) determination was determined by analysis of the NIST standard reference material 1643c, Trace elements in water with the result 19.1 +/- 1.0 mu g Cr(III) l(-1) (certified value 19.0 +/- 0.6 mu g Cr(III) l(-1)). The method was applied to analyse the stability of chromium patches for contact dermatitis testing. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:107 / 113
页数:7
相关论文
共 19 条
[1]   CHROMIUM SPECIATION BY COLUMN CHROMATOGRAPHY USING A DIRECT-CURRENT PLASMA ATOMIC EMISSION SPECTROMETER [J].
AHMAD, S ;
MURTHY, RC ;
CHANDRA, SV .
ANALYST, 1990, 115 (03) :287-289
[2]   INVESTIGATION OF CHROMIUM(III) AND CHROMIUM(VI) SPECIATION IN WATER BY ION CHROMATOGRAPHY WITH CHEMILUMINESCENCE DETECTION [J].
BEERE, HG ;
JONES, P .
ANALYTICA CHIMICA ACTA, 1994, 293 (03) :237-243
[3]   CHROMIUM SPECIATION BY ANION-EXCHANGE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH BOTH INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION SPECTROSCOPIC AND INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRIC DETECTION [J].
BYRDY, FA ;
OLSON, LK ;
VELA, NP ;
CARUSO, JA .
JOURNAL OF CHROMATOGRAPHY A, 1995, 712 (02) :311-320
[4]   FLOW-INJECTION CHEMILUMINESCENCE METHOD FOR THE SELECTIVE DETERMINATION OF CHROMIUM(III) [J].
ESCOBAR, R ;
LIN, QX ;
GUIRAUM, A ;
DELAROSA, FF .
ANALYST, 1993, 118 (06) :643-647
[5]   SIMULTANEOUS DETERMINATION OF CHROMIUM(III) AND CHROMIUM(VI) IN AQUEOUS-SOLUTIONS BY ION CHROMATOGRAPHY AND CHEMILUMINESCENCE DETECTION [J].
GAMMELGAARD, B ;
JONS, O ;
NIELSEN, B .
ANALYST, 1992, 117 (03) :637-640
[6]   PERMEATION OF CHROMIUM SALTS THROUGH HUMAN SKIN INVITRO [J].
GAMMELGAARD, B ;
FULLERTON, A ;
AVNSTORP, C ;
MENNE, T .
CONTACT DERMATITIS, 1992, 27 (05) :302-310
[7]   SIMULTANEOUS DETERMINATION OF CHROMIUM(III) AND CHROMIUM(VI) BY ION CHROMATOGRAPHY WITH INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
INOUE, Y ;
SAKAI, T ;
KUMAGAI, H .
JOURNAL OF CHROMATOGRAPHY A, 1995, 706 (1-2) :127-136
[8]   SIMULTANEOUS DETERMINATION OF CHROMIUM(III) AND CHROMIUM(VI) BY REVERSED-PHASE ION-PAIR HPLC WITH CHROMIUM-SPECIFIC DETECTION [J].
LINTSCHINGER, J ;
KALCHER, K ;
GOSSLER, W ;
KOLBL, G ;
NOVIC, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1995, 351 (07) :604-609
[9]  
Nriagu J.O., 1988, CHROMIUM NATURAL HUM
[10]   SIMULTANEOUS PRECONCENTRATION OF CHROMIUM(III) AND CHROMIUM(VI) PRIOR TO SPECIATION ANALYSIS [J].
OUYANG, GL ;
JEN, JF .
ANALYTICA CHIMICA ACTA, 1993, 279 (02) :329-334