Single frequency bioelectrical impedance is a poor method for determining fat mass in moderately obese women

被引:22
作者
Alvarez, Veronica P.
Dixon, John B. [1 ]
Strauss, Boyd J. G.
Laurie, Cheryl P.
Chaston, Timothy B.
O'Brien, Paul E.
机构
[1] Monash Univ, Alfred Hosp, Monash Med Sch, Australian Ctr Obes Res & Educ, Melbourne, Vic 3004, Australia
[2] Monash Univ, Dept Med, Discipline Nutr & Dietet, Monash Med Ctr, Melbourne, Vic 3004, Australia
关键词
body composition; obese women; weight loss; bioelectrical impedance; fat mass; body mass index; Bland and Altman method; technique comparison;
D O I
10.1007/s11695-007-9032-3
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: The primary aim of weight loss intervention in obesity is the loss of fat mass (FM). Hence, determinations of changes in FM and fat free mass (FFM) during weight loss are of clinical value. The authors compared the clinical utility of SkinFold Thickness (SKF), tetrapolar bioelectrical impedance analysis (BIA) and a body mass index (BMI) based calculation, in determining changes in percentage of fat mass (Delta%FM). Methods: Using dual X-ray absorptiometry (DEXA) measurements of %FM as a standard, BIA, SKF and BMI were compared in 41 moderately obese women (BMI 30-35) before and after significant weight lost (-13.9 +/- 5.8 kg). Results: When measuring fat mass loss, SKF was precise and accurate with a bias of +0.86 +/- 6.16 %, while the BMI-based estimation had a systematic bias of +6.36 +/- 6.04 % (r(2) = 0.791, P<0.001). BIA using the Lukaski formula had a bias of +5.22 % and limits of agreement that approached the magnitude of the measurement (+/- 20.82 %), thus providing no information. In contrast, BIA using the Segal formula had a systematic bias of +7.81 % (r(2) = 0.636, P<0.001) and gave narrower limits of agreement (+/- 8.34 %). Conclusion: For measuring changes in %FM with weight loss, BIA has no clinical value using the Lukaski formula, and using the Segal formula BIA provided no additional information to that given by BMI. We show that BIA instrument variables confound the estimates of %FM achieved by the BMI component of the Lukaski and Segal formulas.
引用
收藏
页码:211 / 221
页数:11
相关论文
共 37 条
[1]   Exercise and diet in obesity treatment: An integrative system dynamics perspective [J].
Abdel-Hamid, TK .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2003, 35 (03) :400-414
[2]  
ALOIA JF, 1995, J NUCL MED, V36, P1392
[3]  
[Anonymous], 2005, IDF CONS WORLDW DEF
[4]  
[Anonymous], [No title captured]
[5]  
Bellanger Tracie Manuel, 2005, J La State Med Soc, V157 Spec No 1, pS42
[6]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]  
BLELANGER TM, 2005, J LA STATE MED SOC, V157, P49
[8]  
Busetto L, 2001, NUTR METAB CARDIOVAS, V11, P195
[9]  
CAMPBELL JN, 1989, TXB PAIN, V2, P22
[10]   Accuracy of bioelectrical impedance spectroscopy in measuring changes in body composition during severe weight loss [J].
Cox-Reijven, PLM ;
van Kreel, B ;
Soeters, PB .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2002, 26 (02) :120-127