Evaluation of LAAPD arrays for high-resolution scintillator matrices readout

被引:5
作者
Kapusta, A [1 ]
Crespo, P
Moszynski, M
Enghardt, W
Szawlowski, M
Zhou, BL
Wolski, D
机构
[1] Soltan Inst Nucl Studies, Otwock, Poland
[2] Gesell Schwerionenforsch mbH, D-64291 Darmstadt, Germany
[3] FZ Rossendorf, D-01314 Dresden, Germany
[4] Adv Photon Inc, Camarillo, CA 93012 USA
关键词
position sensitive avalanche photodiodes; positron emission tomography; scintillation matrices;
D O I
10.1109/TNS.2002.801667
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, we evaluate the performance of recently released avalanche photodiode arrays from Advanced Photonix, Inc., for scintillator matrices readout. The large-area avalanche photodiode (LAAPD) quadrant device is a monolithic 2 x 2 pixel structure with an active area of 5.3 mm(2) per pixel. The device allows stable operation with high gains up to 200 and detection efficiency of 73similar to+/- 10% for 420-nm photons. It is furthermore characterized by a pixel linear fill factor of 90% and low noise equal to 16 electrons equivalent noise charge measured at room temperature. We have measured pixel-to-pixel gain nommiformity smaller than 1.5% and interpixel crosstalk of 1.7% in pulse-mode operation with a device gain of 50. An energy resolution of 12.3 +/- 0.5% was achieved for the 511-keV photopeak from a Na-22 source placed on top of a 2 x 2 x 10 mm(3) LSO crystal coupled to one pixel of a quadrant LAAPD. An array of LSO crystals was later coupled to the device allowing individual crystal identification. A coincidence timing resolution of 1.9 +/- 0.1 ns full-width at half-maximum has been obtained for quadrant pixel with LSO scintillator for the 511-keV peak from a Na-22 source. Finally, we compared the characteristics and readout performance of the LAAPD array with published results of some commercially available avalanche photodiode arrays.
引用
收藏
页码:1693 / 1698
页数:6
相关论文
共 15 条
[1]  
ALLIER CP, 2000, 2000 IEEE NUCL SCI S
[2]   APD arrays and large-area APDs via a new planar process [J].
Farrell, R ;
Shah, K ;
Vanderpuye, K ;
Grazioso, R ;
Myers, R ;
Entine, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 442 (1-3) :171-178
[3]   Avalanche photodiodes in scintillation detection for high resolution PET [J].
Kapusta, M ;
Moszynski, M ;
Balcerzyk, M ;
Lesniewski, K ;
Szawlowski, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (06) :2029-2033
[4]   Energy resolution of scintillation detectors readout with large area avalanche photodiodes and photomultipliers [J].
Moszynski, M ;
Kapusta, M ;
Wolski, D ;
Szawlowski, M ;
Klamra, W .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :472-477
[5]   Comparative study of avalanche photodiodes with different structures in scintillation detection [J].
Moszynski, M ;
Kapusta, M ;
Balcerzyk, M ;
Szawlowski, M ;
Wolski, D ;
Wegrzecka, I ;
Wegrzecki, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (04) :1205-1210
[6]   Large area avalanche photodiodes in X-rays and light detection [J].
Moszynski, M ;
Szawlowski, M ;
Kapusta, M ;
Balcerzyk, M ;
Wolski, D .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (04) :1297-1302
[7]   Subnanosecond timing with large area avalanche photodiodes and LSO scintillator [J].
Moszynski, M ;
Ludziejewski, T ;
Wolski, D ;
Klamra, W ;
Szawlowski, M ;
Kapusta, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1298-1302
[8]   Large Area Avalanche Photodiodes in X-rays and scintillation detection [J].
Moszynski, M ;
Kapusta, M ;
Balcerzyk, M ;
Szawlowski, M ;
Wolski, D .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 442 (1-3) :230-237
[9]   Low energy γ-rays scintillation detection with large area avalanche photodiodes [J].
Moszynski, M ;
Kapusta, M ;
Zalipska, J ;
Balcerzyk, M ;
Wolski, D ;
Szawlowski, M ;
Klamra, W .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (04) :880-885
[10]   Blue enhanced large area avalanche photodiodes in scintillation detection with LSO, YAP and LuAP crystals [J].
Moszynski, M ;
Kapusta, M ;
Wolski, D ;
Szawlowski, M ;
Klamra, W .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :436-442