The assessment of electronic balances for accuracy of mass measurements in the analytical laboratory

被引:16
作者
Gustavo Gonzalez, A. [1 ]
Angeles Herrador, M. [1 ]
机构
[1] Univ Seville, Dept Analyt Chem, E-41012 Seville, Spain
关键词
balance assessment; weighing accuracy; mass uncertainty; balance calibration; buoyancy correction; VALIDATION DATA; UNCERTAINTY; DENSITY; EQUATION;
D O I
10.1007/s00769-006-0214-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The accuracy of mass measurements when using electronic balances is considered within the frame of analytical assays in terms of the uncertainty budget according to the internal quality control routine, the calibration process, the balance specifications, and the weighing scenarios. Buoyancy corrections for both conventional and true mass are fully discussed. The procedure is illustrated with a worked example.
引用
收藏
页码:21 / 29
页数:9
相关论文
共 31 条
[1]  
[Anonymous], 1993, Guide to the Expression of Uncertainty in Measurement
[2]  
[Anonymous], MEASUREMENT GOOD PRA
[3]  
*ASTM, 1993, 898 ASTM E
[4]  
*ASTM, 1991, 617 ASTM E
[5]  
*BIPM IEC IFCC ISO, 1993, INT VOC BAS GEN TERM
[6]  
CLARK JP, 2000, WSRCMS9900958 WEST S
[7]   EQUATION FOR THE DETERMINATION OF THE DENSITY OF MOIST AIR (1981/91) [J].
DAVIS, RS .
METROLOGIA, 1992, 29 (01) :67-70
[8]   Uncertainty of measurement: Approaches and open problems [J].
Desimoni, E ;
Brunetti, B .
ANNALI DI CHIMICA, 2005, 95 (05) :265-274
[9]   Using validation data for ISO measurement uncertainty estimation - Part 1. Principles of an approach using cause and effect analysis [J].
Ellison, SLR ;
Barwick, VJ .
ANALYST, 1998, 123 (06) :1387-1392
[10]  
*EURACHEM CITAC, 2000, EURACHEM CITAC GUID