Radiation temperature scales between the indium and silver points realized at IMGC, NPL and UME using a fixed-point calibration technique

被引:3
作者
Ricolfi, T
Battuello, M
Girard, F
Machin, G
McEvoy, H
Ugur, S
Diril, A
机构
[1] CNR, Ist Metrol G Colonnetti, IMGC, I-10135 Turin, Italy
[2] Natl Phys Lab, Teddington TW11 0LW, Middx, England
[3] TUBITAK, UME, TR-41470 Gebze, Turkey
关键词
temperature; radiation thermometry; fixed points; blackbody radiation; standards and calibration;
D O I
10.1088/0957-0233/13/12/338
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Radiation temperature scales were realized with a precision infrared thermometer at IMGC, NPL and UME using fixed-point blackbodies at the freezing points of indium, tin, zinc, aluminium and silver. From the fixed-point calibration in each laboratory, continuous scales were obtained with a four-coefficient interpolation equation. The differences between the three scales were well within 0.1 degreesC at all temperatures between the indium and silver points. Calibrations of the same thermometer performed at NPL and UME using a more traditional calibration scheme based on the comparison with contact thermometers showed differences of up to more than 0.4 degreesC. This suggested that the fixed-point scheme, that has been used for the first time in an international comparison, should be more reliable for accurate calibrations of precision infrared thermometers.
引用
收藏
页码:2090 / 2093
页数:4
相关论文
共 13 条
[1]  
[Anonymous], 1999, PROC TEMPMEKO 99 7 I
[2]  
[Anonymous], P TEMPEMKO 1999 7 IN
[3]  
Battuello M., 1992, Measurement, V10, P119, DOI 10.1016/0263-2241(92)90004-N
[4]   FIXED-POINT TECHNIQUE FOR APPROXIMATING THE ITS-90 BETWEEN 420-DEGREES-C AND 1085-DEGREES-C WITH AN INFRARED THERMOMETER [J].
BATTUELLO, M ;
LANZA, F ;
RICOLFI, T .
METROLOGIA, 1990, 27 (02) :75-82
[5]  
BATTUELLO M, 2002, IN PRESS TEMPERTURE, V7
[6]  
BATTUELLO M, 2002, IN PRESS TEMPERATURE, V7
[7]  
*EUR COMM, TRAC INFR RAD THERM
[8]  
FISCHER J, 1997, P TEMPMEKO 96 6 INT, P251
[9]   A transportable blackbody furnace for the accurate realization of the fixed points of indium, tin and zinc [J].
Girard, F ;
Ricolfi, T .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1998, 9 (08) :1215-1218
[10]  
MCEVOY HC, 1997, P TEMPMEKO 96, P245