Information in magnetic resonance images: Evaluation of signal, noise and contrast

被引:21
作者
DeWilde, JP [1 ]
Lunt, JA [1 ]
Straughan, K [1 ]
机构
[1] DEPT HLTH,MED DEVICES AGCY,LONDON SE1 6TE,ENGLAND
关键词
contrast-to-noise; correlation; ghosting artefacts; image quality; magnetic resonance imaging; signal-to-noise ratio; Wiener spectrum;
D O I
10.1007/BF02530047
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The assessment of diagnostic image quality for MRI is considered. The assessment of three key image quality determinants is addressed: signal, noise and contrast. There is a distinction between random noise evaluation, for the calculation of the SNR, and structured noise evaluation for the assessment of image artefacts. Specific methods used are correlation techniques and the Wiener spectrum. Contrast is assessed by comparison of experimental data and theoretical predictions, For each assessment, the theory and method of the evaluation strategy are discussed. The discussion is illustrated with analysis results from commercial MR systems. The choice of analysis method and the subsequent derivation of quality indices are shown to be critical in respect of robustness and accuracy.
引用
收藏
页码:259 / 265
页数:7
相关论文
共 21 条
[1]   MAGNETIC-RESONANCE IMAGING - PRESENT STATUS AND FUTURE PERSPECTIVES [J].
BYDDER, GM .
BRITISH JOURNAL OF RADIOLOGY, 1988, 61 (730) :889-897
[2]  
DAINTY JC, 1974, IMAGE SCI, P222
[3]   A SIGNAL-TO-NOISE CALIBRATION PROCEDURE FOR NMR IMAGING-SYSTEMS [J].
EDELSTEIN, WA ;
BOTTOMLEY, PA ;
PFEIFER, LM .
MEDICAL PHYSICS, 1984, 11 (02) :180-185
[4]  
FARRAR TC, 1971, PULSE FOURIER TRANSF, P7
[5]  
HENDRICK R E, 1984, Magnetic Resonance Imaging, V2, P279, DOI 10.1016/0730-725X(84)90193-0
[6]  
KNOWLES RJR, 1988, QUALITY ASSURANCE IM
[8]   MULTI-CENTER TRIAL WITH PROTOCOLS AND PROTOTYPE TEST OBJECTS FOR THE ASSESSMENT OF MRI EQUIPMENT [J].
LERSKI, RA ;
MCROBBIE, DW ;
STRAUGHAN, K ;
WALKER, PM ;
DECERTAINES, JD ;
BERNARD, AM .
MAGNETIC RESONANCE IMAGING, 1988, 6 (02) :201-214
[9]   PRACTICAL ASPECTS OF GHOSTING IN RESISTIVE NUCLEAR-MAGNETIC-RESONANCE IMAGING-SYSTEMS [J].
LERSKI, RA ;
STRAUGHAN, K ;
WILLIAMS, JL .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (07) :721-735
[10]  
LERSKI RA, 1992, RAD MAG, V18, P19