Mobile-bearing total knee arthroplasty - Design factors in minimizing wear

被引:60
作者
Dennis, Douglas A.
Komistek, Richard A.
机构
[1] Colorado Joint Replacement, Denver, CO 80210 USA
[2] Univ Tennessee, Knoxville, TN USA
[3] Rocky Mt Musculoskeletal Res Lab, Denver, CO USA
关键词
D O I
10.1097/01.blo.0000238776.27316.d6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Premature polyethylene wear is a major cause of total knee arthroplasty (TKA) failure. It has been attributed to numerous factors including poor surgical technique, reduced polyethylene thickness, poor locking mechanisms of modular fixed bearing tibial components, gamma irradiation sterilization techniques in the presence of oxygen, and low conformity implant designs. The incidence of implantation of TKA into younger patients who have increased activity requirements and longevity expectations is increasing. This requires continued analysis of design features lessening polyethylene wear. The purpose of this manuscript is to review clinical and basic scientific studies of factors influencing polyethylene wear, focusing on the potential benefits of mobile bearing TKA which potentially reduce long-term polyethylene wear by providing increased implant conformity and reduced polyethylene contact stresses. In vivo kinematic studies have shown self-alignment of the polyethylene bearing with the femoral component typically occurs in rotating platform TKA designs which should hypothetically lessen polyethylene surface stresses, minimize stabilizing post impingement, and increase the potential for enhanced polyethylene longevity.
引用
收藏
页码:70 / 77
页数:8
相关论文
共 69 条
[1]
The influence of an anterior-posterior gliding mobile bearing on range of motion after total knee arthroplasty - A prospective, randomized, double-blinded study [J].
Aigner, C ;
Windhager, R ;
Pechmann, M ;
Rehak, P ;
Engeleke, K .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (10) :2257-2262
[2]
Knee motions during maximum flexion in fixed and mobile-bearing arthroplasties [J].
Banks, S ;
Bellemans, J ;
Nozaki, H ;
Whiteside, LA ;
Harman, M ;
Hodge, WA .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :131-138
[3]
Bargmann LS, 1999, CLIN ORTHOP RELAT R, P49
[4]
THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[5]
THE EFFECT OF CONFORMITY AND PLASTIC THICKNESS ON CONTACT STRESSES IN METAL-BACKED PLASTIC IMPLANTS [J].
BARTEL, DL ;
BURSTEIN, AH ;
TODA, MD ;
EDWARDS, DL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1985, 107 (03) :193-199
[6]
BARTEL DL, 1995, CLIN ORTHOP RELAT R, V317, P76
[7]
BERT JM, 1990, CLIN ORTHOP RELAT R, V254, P211
[8]
A comparison of fixed-bearing and mobile-bearing total knee arthroplasty at a minimum follow-up of 4.5 years [J].
Bhan, S ;
Malhotra, R ;
Kiran, EK ;
Shukla, S ;
Bijjawara, M .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (10) :2290-2296
[9]
Ultra high molecular-weight polyethylene (UHMWPE) in total knee replacement: Fabrication, sterilisation and wear [J].
Blunn, G ;
del Preva, EMB ;
Costa, L ;
Fisher, J ;
Freeman, MAR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (07) :946-949
[10]
BLUNN GW, 1991, CLIN ORTHOP RELAT R, P253