Influence of DNA polymerases on quantitative PCR results using TaqMan™ probe format in the LightCycler™ instrument

被引:20
作者
Kreuzer, KA
Bohn, A
Lass, U
Peters, UR
Schmidt, CA
机构
[1] Humboldt Univ, Med Fak Charite, Innere Med Abt, D-13353 Berlin, Germany
[2] Humboldt Univ, Med Fak Charite, Poliklin MS Hamatol & Onkol, D-13353 Berlin, Germany
关键词
real-time PCR; DNA polymerases; 5 '-exonuclease activity; fluorescent probes;
D O I
10.1006/mcpr.1999.0284
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Real-time fluorescence polymerase chain reaction (PCR) techniques are increasingly used to quantitate target sequences for diagnostic and research purposes. Currently the so called TaqMan(TM) probe chemistry is mostly used as fluorogenic system. This probe format is strictly dependent on the 5'-exonuclease activity of DNA polymerase as fragmentation of the probe during the reaction is essential for this assay. Based on our experience that dramatic differences in quantitative PCR results may be due to different DNA polymerases we performed a detailed comparison of 15 enzymes. We found that clear differences exist between polymerases of different manufacturers. Thus, three out of seven polymerases which were declared to posses 5'-exonuclease activity appeared to be completely unsuitable for this method while the remaining had significantly different reaction efficiencies. We conclude that different DNA polymerases may determine the entire analytical performance of TaqMan(TM) assays suggesting that DNA polymerase testing is of special importance when this probe format is used. (C) 2000 Academic Press.
引用
收藏
页码:57 / 60
页数:4
相关论文
共 4 条
[1]  
LIVAK KJ, 1995, PCR METH APPL, V4, P357
[2]   STRUCTURE-SPECIFIC ENDONUCLEOLYTIC CLEAVAGE OF NUCLEIC-ACIDS BY EUBACTERIAL DNA-POLYMERASES [J].
LYAMICHEV, V ;
BROW, MAD ;
DAHLBERG, JE .
SCIENCE, 1993, 260 (5109) :778-783
[3]  
RYCHLIK W, 1995, BIOTECHNIQUES, V23, P714
[4]   The LightCycler(TM) a microvolume multisample fluorimeter with rapid temperature control [J].
Wittwer, CT ;
Ririe, KM ;
Andrew, RV ;
David, DA ;
Gundry, RA ;
Balis, UJ .