Failure analysis of a Ti6Al4V cementless HIP prosthesis

被引:48
作者
Chao, Jesus [1 ]
Lopez, Victor [1 ]
机构
[1] CSIC, Ctr Nacl Invest Met, E-28040 Madrid, Spain
关键词
hip prosthesis; fatigue failure; Ti6A14V; microstructure; notch effect;
D O I
10.1016/j.engfailanal.2006.11.003
中图分类号
TH [机械、仪表工业];
学科分类号
0802 [机械工程];
摘要
This work investigates the causes of in vivo fracture of a cementless hip prosthesis model manufactured with Ti6A14V alloy. To this end an assessment of geometric characteristics, a fractographic analysis of the fracture surface, and a microstructural study of the stem material were carried out. The results indicate that the fracture was initiated at the junction radius of the neck with the collared device, with around 90% of the fracture surface being due to a fatigue mechanism, In order to assess the impact of the material's microstructure on its mechanical properties, two heat treatments, yielding a fine and a coarse microstructure, respectively, were performed on a 14 mm diameter bar. The notch effect on fatigue strength was evaluated by comparison of the results of polished specimens and specimens having a notch whose geometry produced a similar stress concentration factor to that which caused the fracture in the implant. It is shown that irrespective of the material's microstructure, a notch greatly reduces the fatigue strength of this material. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:822 / 830
页数:9
相关论文
共 23 条
[1]
American Society for Testing andMaterials, 2002, F13602A ASTM
[2]
[Anonymous], 1996, 58323 ISO
[3]
*ASTM, 2003, E801 ASTM, P61
[4]
*ASTM, 2003, E2302 ASTM, P147
[5]
BARTLO LJ, 1969, ASTM STP, V459, P144, DOI DOI 10.1016/J.MATDES.2012.10.059
[6]
CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES [J].
COLLIER, JP ;
SURPRENANT, VA ;
JENSEN, RE ;
MAYOR, MB ;
SURPRENANT, HP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04) :511-517
[7]
FATIGUE FAILURE OF NONCEMENTED POROUS-COATED IMPLANTS - A RETRIEVAL STUDY [J].
COOK, SD ;
THOMAS, KA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (01) :20-24
[8]
THE EFFECT OF POST-SINTERING HEAT-TREATMENTS ON THE FATIGUE PROPERTIES OF POROUS COATED TI-6AL-4V ALLOY [J].
COOK, SD ;
THONGPREDA, N ;
ANDERSON, RC ;
HADDAD, RJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (04) :287-302
[9]
COOK SD, 1984, J BIOMED MATER RES, V18, P312
[10]
Frost NE, 1974, OXFORD ENG SCI SERIE, P130