The university of Florida frameless high-precision stereotactic radiotherapy system

被引:131
作者
Bova, FJ
Buatti, JM
Friedman, WA
Mendenhall, WM
Yang, CC
Liu, C
机构
[1] Department of Radiation Oncology, Univ. of Florida Hlth. Sci. Center, Gainesville, FL
[2] Department of Radiation Oncology, Univ. of Florida Hlth. Sci. Center, Gainesville, FL 32610-0385
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1997年 / 38卷 / 04期
关键词
high precision radiotherapy; immobilization; stereotactic radiotherapy; radiosurgery;
D O I
10.1016/S0360-3016(97)00055-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To develop and test a system for high precision fractionated stereotactic radiotherapy that separates immobilization and localization devices. Methods and Materials: Patient localization is achieved through detection and digital registration of an independent bite plate system. The bite plate is made and linked to a set of six infrared light emitting diodes (IRLEDs). These IRLEDs are detected by an infrared camera system that identifies the position of each IRLED within 0.1 to 0.15 mm. Calibration of the camera system defines isocenter and translational X, Y, and Z axes of the stereotactic radiosurgery subsystem and thereby digitally defines the virtual treatment room space in a computer linked to the camera system. Positions of the bite plate's IRLEDs are processed digitally using a computer algorithm so that positional differences between an actual bite plate position and a desired position can be resolved within 0.1 mm of translation (X, Y, and Z distance) and 0.1 degree of rotation. Furthermore, bite plate misalignment can be displayed digitally in real time with translational (x, y, and z) and rotational (roll, pitch, and yawl parameters for an actual bite plate position. Immobilization is achieved by a custom head mold and thermal plastic mask linked by hook-and-loop fastener tape. The head holder system permits rotational and translational movements for daily treatment positioning based on the bite plate localization system. Initial testing of the localization system was performed on 20 patients treated with radiosurgery. The system was used to treat 11 patients with fractionated stereotactic radiotherapy. Results: Assessment of bite plate localization in radiosurgery patients revealed that the patient's bite plate could be positioned and repositioned within 0.5 +/- 0.3 mm (standard deviation). After adjustments, the first 11 patients were treated with the bite plate repositioning error reduced to 0.2 +/- 0.1 mm. Conclusions: High precision stereotactic radiotherapy can be delivered using separate localization and immobilization systems. Treatment setup and delivery can be accomplished in 15 min or less. Advantages compared with standard systems require further study. (C) 1997 Elsevier Science Inc.
引用
收藏
页码:875 / 882
页数:8
相关论文
共 28 条
[1]  
BOVA FJ, 1992, MED PROG TECHNOL, V18, P239
[2]   FRACTIONATED RADIOTHERAPY OF SMALL INOPERABLE LESIONS OF THE BRAIN USING A NONINVASIVE STEREOTAXIC FRAME [J].
DELANNES, M ;
DALY, NJ ;
BONNET, J ;
SABATIER, J ;
TREMOULET, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03) :749-755
[3]  
DICKSON DR, 1994, 42 ANN M RAD RES SOC
[4]   STEREOTAXIC RADIOTHERAPY FOR PEDIATRIC AND ADULT BRAIN-TUMORS - PRELIMINARY-REPORT [J].
DUNBAR, SF ;
TARBELL, NJ ;
KOOY, HM ;
ALEXANDER, E ;
BLACK, PM ;
BARNES, PD ;
GOUMNEROVA, L ;
SCOTT, RM ;
POMEROY, SL ;
LAVALLY, B ;
SALLAN, SE ;
LOEFFLER, JS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (03) :531-539
[5]  
Friedman W A, 1990, Neurosurg Clin N Am, V1, P991
[6]   THE UNIVERSITY-OF-FLORIDA RADIOSURGERY SYSTEM [J].
FRIEDMAN, WA ;
BOVA, FJ .
SURGICAL NEUROLOGY, 1989, 32 (05) :334-342
[7]   FRACTIONATED STEREOTAXICALLY GUIDED RADIOTHERAPY OF HEAD AND NECK TUMORS - A REPORT ON CLINICAL USE OF A NEW SYSTEM IN 195 CASES [J].
GADEMANN, G ;
SCHLEGEL, W ;
DEBUS, J ;
SCHAD, L ;
BORTFELD, T ;
HOVER, KH ;
LORENZ, WJ ;
WANNENMACHER, M .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :205-213
[8]   RELOCATABLE FRAME FOR STEREOTAXIC EXTERNAL BEAM RADIOTHERAPY [J].
GILL, SS ;
THOMAS, DGT ;
WARRINGTON, AP ;
BRADA, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 20 (03) :599-603
[9]   MULTIDIMENSIONAL TREATMENT PLANNING .2. BEAM EYE-VIEW, BACK PROJECTION, AND PROJECTION THROUGH CT SECTIONS [J].
GOITEIN, M ;
ABRAMS, M ;
ROWELL, D ;
POLLARI, H ;
WILES, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (06) :789-797
[10]   CALCULATION OF THE UNCERTAINTY IN THE DOSE DELIVERED DURING RADIATION-THERAPY [J].
GOITEIN, M .
MEDICAL PHYSICS, 1985, 12 (05) :608-612