Safety of combined heat and moisture exchanger filters in long-term mechanical ventilation

被引:58
作者
Hurni, JM
Feihl, F
Lazor, R
Leuenberger, P
Perret, C
机构
[1] CHU VAUDOIS, INST PHYSIOPATHOL CLIN, CH-1011 LAUSANNE, SWITZERLAND
[2] CHU VAUDOIS, DIV PNEUMOL, CH-1011 LAUSANNE, SWITZERLAND
关键词
airway humidification; heat and moisture exchanger; intensive care; mechanical ventilation;
D O I
10.1378/chest.111.3.686
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study objective: To evaluate the safety of a combined heat and moisture exchanger filter (HMEF) for the conditioning of inspired gas in long-term mechanical ventilation (MV). Design: Randomized controlled trial. Setting: Medical ICU in a large teaching hospital. Patients: One hundred fifteen consecutive patients who required greater than or equal to 48 h of MV. Interventions: Patients were randomized at intubation time (day 1) to receive inspired gas conditioned either by a water-bath humidifier heated at 32 degrees C (HWBH) or by an HMEF (Hygroster; DAR; Mirandola, Italy). Measurements and main results: The two study groups were comparable in terms of primary pathologic condition at the time of hospital admission, disease severity as measured by the Simplified Acute Physiology Score, and ICU mortality. They did not differ with respect to ventilator days per patient (mean +/- SD: HMEF, 7.6 +/- 6.5; HWBH, 7.8 +/- 5.8), incidence of endotracheal tube obstruction (HMEF, 0/59; HWBH, 1/56), and incidence of hypothermic episodes (HMEF, five; HWBH, two). In 41 patients receiving MV for greater than or equal to 5 days, the morphologic integrity of respiratory epithelium was evaluated on day 1 and day 5, using a cytologic examination of tracheal aspirate smears. The state of ciliated epithelium was scored on a scale from 0 (poorest integrity) to 1,200 (maximum integrity), according to a well-described method. In both patient groups, the scores slightly but significantly decreased from day 1 to day 5 (mean +/- SD: HWBH, from 787 +/- 104 to 745 +/- 88; HMEF, from 813 +/- 79 to 739 +/- 62; p < 0.01 for both groups); there were no statistically significant differences between groups. Conclusions: These data indicate acceptable safety of HMEFs of the type used in the present study for long-term mechanical ventilation.
引用
收藏
页码:686 / 691
页数:6
相关论文
共 22 条
[1]   HUMIDIFICATION IN THE INTENSIVE-CARE UNIT - PROSPECTIVE-STUDY OF A NEW PROTOCOL UTILIZING HEATED HUMIDIFICATION AND A HYGROSCOPIC CONDENSER HUMIDIFIER [J].
BRANSON, RD ;
DAVIS, K ;
CAMPBELL, RS ;
JOHNSON, DJ ;
POREMBKA, DT .
CHEST, 1993, 104 (06) :1800-1805
[2]  
CHALON J, 1980, B NEW YORK ACAD MED, V56, P314
[3]   EFFECTS OF DRY ANESTHETIC GASES ON TRACHEOBRONCHIAL CILIATED EPITHELIUM [J].
CHALON, J ;
MALEBRANCHE, J ;
LOEW, DAY .
ANESTHESIOLOGY, 1972, 37 (03) :338-+
[4]   HUMIDITY AND THE ANESTHETIZED PATIENT [J].
CHALON, J ;
PATEL, C ;
ALI, M ;
RAMANATHAN, S ;
CAPAN, L ;
TANG, CK ;
TURNDORF, H .
ANESTHESIOLOGY, 1979, 50 (03) :195-198
[5]   ENDOTRACHEAL-TUBE OCCLUSION ASSOCIATED WITH THE USE OF HEAT AND MOISTURE EXCHANGERS IN THE INTENSIVE-CARE UNIT [J].
COHEN, IL ;
WEINBERG, PF ;
FEIN, IA ;
ROWINSKI, GS .
CRITICAL CARE MEDICINE, 1988, 16 (03) :277-279
[6]  
DREYFUSS D, 1995, AM J RESP CRIT CARE, V151, P986
[7]   HEAT AND MOISTURE EXCHANGERS AND THE BODY-TEMPERATURE - A PEROPERATIVE STUDY [J].
ECKERBOM, B ;
LINDHOLM, CE .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1990, 34 (07) :538-542
[8]  
FEIHL F, 1992, Acta Anaesthesiologica Italica, V43, P57
[9]  
Hess D, 1994, PRINCIPLES PRACTICE, P1039
[10]   AN EVALUATION OF THE HEAT AND MOISTURE EXCHANGE PERFORMANCE OF 4 VENTILATOR CIRCUIT FILTERS [J].
JACKSON, C ;
WEBB, AR .
INTENSIVE CARE MEDICINE, 1992, 18 (05) :264-268