Improved dc SQUID read-out electronics with low 1/f noise preamplifier

被引:37
作者
Drung, D
机构
[1] Phys.-Technische Bundesanstalt, D-10587 Berlin
关键词
D O I
10.1063/1.1148348
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An improved read-out electronics without flux modulation for de superconducting quantum interference devices (de SQUIDs) is presented where the preamplifier is directly coupled to the SQUID. Compared to our previous designs, the 1/f voltage and current noise levels of the preamplifier were reduced by about a factor of 4 (rms) by using bipolar matched transistors instead of operational amplifiers at the input. Three types of flux-locked-loop electronics were developed and successfully tested: a low-power (120 mW) version for large multichannel arrays, a high-speed version with a 15 MHz bandwidth and a very fast (300 ns) reset mode, and a low-noise (0.44 nV/root<(Hz)over bar>) version with a fast and precise bias reversal circuit. The flux-locked-loop electronics are compatible with each other, so that only one type of low-power (180 mW) control unit is required to operate them. All essential specifications of practical SQUID systems (noise, dynamic behavior, and temperature stability) were determined by using, Liquid-helium and liquid-nitrogen cooled SQUIDs, respectively. (C) 1997 American Institute of Physics. [S0034-6748(97)02511-2].
引用
收藏
页码:4066 / 4074
页数:9
相关论文
共 20 条
[1]   TUNNEL JUNCTION DC SQUID - FABRICATION, OPERATION, AND PERFORMANCE [J].
CLARKE, J ;
GOUBAU, WM ;
KETCHEN, MB .
JOURNAL OF LOW TEMPERATURE PHYSICS, 1976, 25 (1-2) :99-144
[2]   THEORY FOR THE MULTILOOP DC SUPERCONDUCTING QUANTUM INTERFERENCE DEVICE MAGNETOMETER AND EXPERIMENTAL-VERIFICATION [J].
DRUNG, D ;
KNAPPE, S ;
KOCH, H .
JOURNAL OF APPLIED PHYSICS, 1995, 77 (08) :4088-4098
[3]   THE PTB 83-SQUID SYSTEM FOR BIOMAGNETIC APPLICATIONS IN A CLINIC [J].
DRUNG, D .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1995, 5 (02) :2112-2117
[4]   AN INTEGRATED DC SQUID MAGNETOMETER WITH VARIABLE ADDITIONAL POSITIVE FEEDBACK [J].
DRUNG, D ;
KOCH, H .
SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 1994, 7 (05) :242-245
[5]  
Drung D, 1996, NATO ADV SCI I E-APP, V329, P63
[6]   LOW-FREQUENCY NOISE IN LOW-T-C MULTILOOP MAGNETOMETERS WITH ADDITIONAL POSITIVE FEEDBACK [J].
DRUNG, D .
APPLIED PHYSICS LETTERS, 1995, 67 (10) :1474-1476
[7]   AN ELECTRONIC SECOND-ORDER GRADIOMETER FOR BIOMAGNETIC APPLICATIONS IN CLINICAL SHIELDED ROOMS [J].
Drung, D. ;
Koch, H. .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1993, 3 (01) :2594-2597
[8]   A 2-channel wideband SQUID system for high-frequency geophysical applications [J].
Drung, D ;
Radic, T ;
Matz, H ;
Koch, H ;
Knappe, S ;
Menkel, S ;
Burkhardt, H .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1997, 7 (02) :3283-3286
[9]   A 5-MHZ BANDWIDTH SQUID MAGNETOMETER WITH ADDITIONAL POSITIVE FEEDBACK [J].
DRUNG, D ;
MATZ, H ;
KOCH, H .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (04) :3008-3015
[10]   INTEGRATED DC SQUID MAGNETOMETER WITH HIGH DV/DB [J].
DRUNG, D ;
CANTOR, R ;
PETERS, M ;
RYHANEN, T ;
KOCH, H .
IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) :3001-3004