The linear electrode array: a useful tool with many applications

被引:217
作者
Merletti, R [1 ]
Farina, D [1 ]
Gazzoni, M [1 ]
机构
[1] Politecn Torino, Dipartimento Elettron, Ctr Bioingn, I-10129 Turin, Italy
关键词
electromyography; surface EMG; electrode arrays; motor unit; innervation zone; muscle fiber conduction velocity;
D O I
10.1016/S1050-6411(02)00082-2
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
In this review we describe the basic principles of operation of linear electrode arrays for the detection of surface EMG signals, together with their most relevant current applications. A linear array of electrodes is a system which detects surface EMG signals in a number of points located along a line. A spatial filter is usually placed in each point for signal detection, so that the recording of EMG signals with linear arrays corresponds to the sampling in one spatial direction of a spatially filtered version of the potential distribution over the skin. Linear arrays provide indications on motor unit (MU) anatomical properties, such as the locations of the innervation zones and tendons, and the fiber length. Such systems allow the investigation of the properties of the volume conductor and its effect on surface detected signals. Moreover, linear arrays allow to estimate muscle fiber conduction velocity with a very low standard deviation of estimation (of the order of 0.1-0.2 m/s), thus providing reliable indications on muscle fiber membrane properties and their changes in time (for example with fatigue or during treatment). Conduction velocity can be estimated from a signal epoch (global estimate) or at the single MU level. In the latter case, MU action potentials are identified from the interference EMG signals and conduction velocity is estimated for each detected potential. In this way it is possible, in certain conditions, to investigate single MU control and conduction properties with a completely non-invasive approach. Linear arrays provide valuable information on the neuromuscular system properties and appear to be promising tools for applied studies and clinical research. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:37 / 47
页数:11
相关论文
共 63 条
[1]  
Agnivesh C. R., 2002, TOXICOLOGY AYURVEDIC, P23
[2]   MUSCLE-FIBER CONDUCTION-VELOCITY IN MOTOR UNITS OF THE HUMAN ANTERIOR TIBIAL MUSCLE - A NEW SIZE PRINCIPLE PARAMETER [J].
ANDREASSEN, S ;
ARENDTNIELSEN, L .
JOURNAL OF PHYSIOLOGY-LONDON, 1987, 391 :561-571
[3]   MEASUREMENT OF MUSCLE-FIBER CONDUCTION-VELOCITY IN HUMANS - TECHNIQUES AND APPLICATIONS [J].
ARENDTNIELSEN, L ;
ZWARTS, M .
JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 1989, 6 (02) :173-190
[4]  
BLOK J, 1999, CLIN NEUROPHYSIOL, V110, pS90
[5]   A high-density multichannel surface electromyography system for the characterization of single motor units [J].
Blok, JH ;
van Dijk, JP ;
Drost, G ;
Zwarts, MJ ;
Stegeman, DF .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (04) :1887-1897
[6]   MYOELECTRIC SIGNAL CONDUCTION-VELOCITY AND SPECTRAL PARAMETERS - INFLUENCE OF FORCE AND TIME [J].
BROMAN, H ;
BILOTTO, G ;
DE LUCA, CJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1985, 58 (05) :1428-1437
[7]  
BROMAN H, 1985, IEEE T BIOMED ENG, V32, P311
[8]  
Chisari C, 1998, MUSCLE NERVE, V21, P543
[9]  
Dimitrov G V, 2002, Electromyogr Clin Neurophysiol, V42, P231
[10]  
DIMITROVA N, 1974, Electromyography and Clinical Neurophysiology, V14, P53