Multiple site analytical evaluation of a portable blood gas/electrolyte analyzer for point of care testing

被引:8
作者
Chance, JJ
Li, DJ
Sokoll, LJ
Silberman, MA
Engelstad, ME
Nichols, JH
Liu, XH
Mohammad, AA
Petersen, JR
Okorodudu, AO
机构
[1] Johns Hopkins Hosp, Dept Pathol, Div Clin Chem, Baltimore, MD 21287 USA
[2] Univ Texas, Med Branch, Dept Pathol, Div Clin Chem, Galveston, TX 77550 USA
关键词
critical care medicine; point of care testing; blood gases; electrolytes; portable; tonometry; cost; correlation;
D O I
10.1097/00003246-200006000-00069
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate the analytical performance of the SenDx 100 portable blood gas and electrolyte analyzer (SenDx Medical, Carlsbad, CA). Design: Accuracy was evaluated by correlation of whole blood patient samples with the Nova Stat Profile 5 (Nova Biomedical, Waltham, MA) and the Ciba Coming 865 (Chiron Diagnostics, Medford, MA). Precision was evaluated using quality control materials (RNA Medical, Acton, MA). Setting: Critical care laboratories and operating rooms in two institutions. Measurements and Main Results:Precision studies performed at three different concentration levels for each analyte demonstrated intra-assay precision of less than or equal to 2.5% coefficient of variation and interassay precision of less than or equal to 4.0% coefficient of variation in all cases. Analysis of patient specimens in general showed good to excellent correlation to reference analyzers. Regression Variables are tabulated. Conclusions: The SenDx 100 portable blood gas and electrolyte analyzer is a simple and easy to use analyzer demonstrating acceptable performance compared with reference methods.
引用
收藏
页码:2081 / 2085
页数:5
相关论文
共 8 条
[1]  
BURKE MD, 1993, AM J CLIN PATHOL, V100, P89
[2]  
*COLL AM PATH, 1998, CRIT CAR AQ BLOOD GA
[3]   Evaluation of i-STAT portable clinical analyzer in a hemodialysis unit [J].
Gault, MH ;
Harding, CE .
CLINICAL BIOCHEMISTRY, 1996, 29 (02) :117-124
[4]  
HERR DM, 1995, AM J CLIN PATHOL, V104, pS54
[5]  
JACOBS E, 1993, CLIN CHEM, V39, P1890
[6]  
Lamb L S Jr, 1995, Am J Crit Care, V4, P429
[7]   Evaluation of i-STAT portable clinical analyzer in a neonatal and pediatric intensive care unit [J].
Murthy, JN ;
Hicks, JM ;
Soldin, SJ .
CLINICAL BIOCHEMISTRY, 1997, 30 (05) :385-389
[8]  
WONG RJ, 1994, CLIN CHEM, V40, P124