Assessment of a novel implantable artificial anal sphincter

被引:28
作者
Hajivassiliou, CA
Carter, KB
Finlay, IG
机构
[1] GLASGOW ROYAL INFIRM,DEPT COLOPROCTOL,GLASGOW G4 0SF,LANARK,SCOTLAND
[2] GLASGOW ROYAL INFIRM,DEPT MED PHYS,GLASGOW G4 0SF,LANARK,SCOTLAND
关键词
artificial anal sphincter; artificial implants; fecal incontinence; prosthesis;
D O I
10.1007/BF02140902
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
PURPOSE: The aim of the study was to test a new implantable artificial anal sphincter in the porcine model. METHOD: The artificial sphincter, which includes an inflatable expander that compresses and flattens the bowel against a pillow, was implanted in 16 animals and studied for periods of up to 20 weeks. The anal sphincters were destroyed, and the efficacy of the device in rendering the animals continent was studied. RESULTS: Of the 11 animals in which the artificial sphincter was regularly closed, 8 completed the study and were continent during 85 percent of activation times. There was no evidence of ischemic injury. Major complications were related only to failure of the control pumps of the device. CONCLUSION: This study suggests that this neosphincter produces fecal continence without intestinal ischemia. At present reliability is limited only by the performance of the pump.
引用
收藏
页码:711 / 717
页数:7
相关论文
共 24 条
[1]  
Bartolo D, 1987, COLOPROCTOLOGY, V9, P103
[2]   THE LONG-TERM RESULTS OF ARTIFICIAL SPHINCTERS IN CHILDREN [J].
BOSCO, PJ ;
BAUER, SB ;
COLODNY, AH ;
MANDELL, J ;
RETIK, AB .
JOURNAL OF UROLOGY, 1991, 146 (02) :396-399
[3]   LONG-TERM RESULTS OF POSTANAL REPAIR FOR NEUROGENIC FECAL INCONTINENCE [J].
CARRARO, PS ;
KAMM, MA ;
NICHOLLS, RJ .
BRITISH JOURNAL OF SURGERY, 1994, 81 (01) :140-144
[4]   TREATMENT OF ANAL INCONTINENCE BY AN IMPLANTABLE PROSTHETIC ANAL-SPHINCTER [J].
CHRISTIANSEN, J ;
SPARSO, B .
ANNALS OF SURGERY, 1992, 215 (04) :383-386
[5]   PROSTHETIC SPHINCTER FOR GASTROINTESTINAL-TRACT [J].
DELANEY, J ;
BROADIE, T ;
TIMM, G ;
BRADLEY, W .
JOURNAL OF SURGICAL RESEARCH, 1974, 16 (03) :204-209
[6]   CIVILIAN AND WAR INJURIES OF THE PERINEUM AND ANAL SPHINCTERS [J].
ENGEL, AF ;
KAMM, MA ;
HAWLEY, PR .
BRITISH JOURNAL OF SURGERY, 1994, 81 (07) :1069-1073
[7]   THE USE OF THE AMS-AS800 ARTIFICIAL SPHINCTER FOR CONTINENT URINARY-DIVERSION .1. INVESTIGATIONS, INCLUDING PRESSURE-FLOW STUDIES, USING RABBIT INTESTINAL LOOPS [J].
ENGELMANN, UH ;
FELDERMAN, TP ;
SCOTT, FB .
JOURNAL OF UROLOGY, 1985, 134 (01) :183-186
[8]   EVALUATION OF AMS-800 ARTIFICIAL SPHINCTER FOR CONTINENT URINARY-DIVERSION USING INTESTINAL LOOPS [J].
ENGELMANN, UH ;
FELDERMAN, TP ;
SCOTT, FB .
UROLOGY, 1985, 25 (06) :620-621
[9]  
Ersek R A, 1991, J Long Term Eff Med Implants, V1, P155
[10]   PHYSIOLOGICAL AND HISTOCHEMICAL ADAPTATION OF THE ELECTRICALLY STIMULATED GRACILIS MUSCLE TO NEOANAL SPHINCTER FUNCTION [J].
GEORGE, BD ;
WILLIAMS, NS ;
PATEL, J ;
SWASH, M ;
WATKINS, ES .
BRITISH JOURNAL OF SURGERY, 1993, 80 (10) :1342-1346