Laparoscopic partial kidney ablation with high intensity focused ultrasound

被引:39
作者
Paterson, RF [1 ]
Barret, E
Siqueira, TM
Gardner, TA
Tavakkoli, J
Rao, VV
Sanghvi, NT
Cheng, L
Shalhav, AL
机构
[1] Indiana Univ, Sch Med, Dept Urol, Inst Kidney Stone Dis,Methodist Hosp, Indianapolis, IN 46202 USA
[2] Indiana Univ, Sch Med, Dept Pathol, Indianapolis, IN USA
[3] Hop Pitie, Serv Urol, F-75651 Paris, France
[4] Focus Surg Inc, Indianapolis, IN USA
关键词
kidney; laparoscopy; ultrasonic therapy; equipment and supplies; swine; miniature;
D O I
10.1016/S0022-5347(05)64124-9
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: High intensity focused ultrasound has been performed for transrectal and extracorporeal thermal ablation of tissues. We developed and tested a laparoscopic probe that allows real-time ultrasound imaging during partial renal ablation using high intensity focused ultrasound. Methods: A Sonablate 200 (Focus Surgery, Indianapolis, Indiana) high intensity focused ultrasound system with a modified 18 mm. laparoscopic probe was used in all experiments. In 13 Yucatan mini-pigs a 5Fr ureteral catheter was inserted into the renal pelvis and 10 cc air were instilled into the collecting system. The kidney was laparoscopically dissected, the high intensity focused ultrasound probe was inserted through a 33 mm. laparoscopic port and the targeted renal pole was treated. Results: Renal lesions were created in 12 of 13 treated kidneys under real-time ultrasound visualization. Median operative time was 180 minutes, average high intensity focused ultrasound activation time was 18.3 minutes and lesion size was 23 X 17 X 11 mm. At 4 and 14 days 4 (acute group) and 6 (subacute group) animals were available for renal functional and anatomical evaluation, respectively. No difference in renal function was seen in treated and untreated kidneys. Pathological examination at 14 days revealed homogenous and complete tissue necrosis throughout the whole volume of the lesion with sharp demarcation from adjacent normal tissue. Conclusions: We were able to refine a probe for laparoscopic high intensity focused ultrasound delivery capable of simultaneous ultrasound imaging. Partial renal ablation using this probe is feasible and safe, and resulted in homogenous, complete and reproducible lesions.
引用
收藏
页码:347 / 351
页数:5
相关论文
共 20 条
[1]   High-intensity focused ultrasound ablation of rabbit kidney tumors [J].
Adams, JB ;
Moore, RG ;
Anderson, JH ;
Strandberg, JD ;
Marshall, FF ;
Davoussi, LR .
JOURNAL OF ENDOUROLOGY, 1996, 10 (01) :71-75
[2]  
CHAPELON JY, 1992, EUR UROL, V22, P147
[3]   Influence of ablated tissue on the formation of high-intensity focused ultrasound lesions [J].
Chen, LL ;
terHaar, G ;
Hill, CR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (06) :921-931
[4]   Laparoscopic and percutaneous radiofrequency ablation of the kidney: Acute and chronic porcine study [J].
Gill, IS ;
Hsu, THS ;
Fox, RL ;
Matamoros, A ;
Miller, CD ;
LeVeen, RF ;
Grune, MT ;
Sung, GT ;
Fidler, ME .
UROLOGY, 2000, 56 (02) :197-200
[5]   Laparoscopic renal cryoablation: Initial clinical series [J].
Gill, IS ;
Novick, AC ;
Soble, JJ ;
Sung, GT ;
Remer, EM ;
Hale, J ;
O'Malley, CM .
UROLOGY, 1998, 52 (04) :543-551
[6]  
Herts B R, 1995, Semin Urol Oncol, V13, P254
[7]   High intensity focused ultrasound as noninvasive therapy for multilocal renal cell carcinoma:: Case study and review of the literature [J].
Köhrmann, KU ;
Michel, MS ;
Gaa, J ;
Marlinghaus, E ;
Alken, P .
JOURNAL OF UROLOGY, 2002, 167 (06) :2397-2403
[8]   Laparoscopic partial nephrectomy and wedge resection for the treatment of renal malignancy [J].
Kozlowski, PM ;
Winfield, HN .
JOURNAL OF ENDOUROLOGY, 2001, 15 (04) :369-374
[9]  
Lindberg Guy, 2000, Journal of Endourology, V14, pA1
[10]  
McDougall E M, 1998, JSLS, V2, P15