A real-time microprocessor QRS detector system with a 1-ms timing accuracy for the measurement of ambulatory HRV

被引:194
作者
Ruha, A
Sallinen, S
Nissila, S
机构
[1] University of Oulu, Department of Electrical Engineering
[2] Varian-Dosetek Oy, Espoo FIN-02130, Tietajantie
关键词
biomedical monitoring; electrocardiography; matched filters; signal detection; timing jitter;
D O I
10.1109/10.554762
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The design, test methods, and results of an ambulatory QRS detector are presented, The device is intended for the accurate measurement of heart rate variability (HRV) and reliable QRS detection in both ambulatory and clinical use, The aim of the design work was to achieve high QRS detection performance in terms of timing accuracy and reliability, without compromising the size and power consumption of the de,ice. The complete monitor system consists of a host computer and the detector unit, The detector de,ice is constructed of a commonly available digital signal processing (DSP) microprocessor and other components, The QRS detection algorithm uses optimized prefiltering in conjunction with a matched filter and dual edge threshold detection, The purpose of the prefiltering is to attenuate various noise components in order to achieve improved detection reliability, The matched filter further improves signal-to-noise ratio (SNR) and symmetries the QRS complex for the threshold detection, which is essential in order to achieve the desired performance. The decision for detection is made in real-time and no search-bark method is employed, The host computer is used to configure the detector unit, which includes the setting of the matched filter impulse response, and in the retrieval and postprocessing of the measurement results, The QRS detection timing accuracy and detection reliability of the detector system was tested with an artificially generated electrocardiogram (EGG) signal corrupted with various noise types and a timing standard deviation of less than 1 ms was achieved with most noise types and levels similar to those encountered in real measurements, A QRS detection error rate (ER) of 0.1 and 2.2% was achieved with records 103 and 105 from the MIT-BM Arrhythmia database, respectively.
引用
收藏
页码:159 / 167
页数:9
相关论文
共 14 条
[1]  
*AAMI, 1981, STAND CARD MON HEART
[2]  
BARBARO V, 1991, MED BIOL ENG COMPUT, V21, P129
[3]   A COMPARISON OF THE NOISE SENSITIVITY OF 9 QRS DETECTION ALGORITHMS [J].
FRIESEN, GM ;
JANNETT, TC ;
JADALLAH, MA ;
YATES, SL ;
QUINT, SR ;
NAGLE, HT .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (01) :85-98
[4]   QUANTITATIVE INVESTIGATION OF QRS DETECTION RULES USING THE MIT/BIH ARRHYTHMIA DATABASE [J].
HAMILTON, PS ;
TOMPKINS, WJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (12) :1157-1165
[5]   NEW SOFTWARE QRS DETECTOR ALGORITHM SUITABLE FOR REAL-TIME APPLICATIONS WITH LOW SIGNAL-TO-NOISE RATIOS [J].
LINDECRANTZ, KG ;
LILJA, H .
JOURNAL OF BIOMEDICAL ENGINEERING, 1988, 10 (03) :280-284
[6]   PROGNOSTIC VALUE OF HEART-RATE VARIABILITY AFTER MYOCARDIAL-INFARCTION - A COMPARISON OF DIFFERENT DATA-PROCESSING METHODS [J].
MALIK, M ;
CRIPPS, T ;
FARRELL, T ;
CAMM, AJ .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1989, 27 (06) :603-611
[7]   A REAL-TIME QRS DETECTION ALGORITHM [J].
PAN, J ;
TOMPKINS, WJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1985, 32 (03) :230-236
[8]  
PEEBLES PZ, 1980, PROBABILITY RANDOM V, P257
[9]  
SCHWARZ M, 1970, INFORMATION TRANSMIS, P577
[10]   ESTIMATION OF QRS COMPLEX POWER SPECTRA FOR DESIGN OF A QRS FILTER [J].
THAKOR, NV ;
WEBSTER, JG ;
TOMPKINS, WJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (11) :702-706