摘要
Objective: To establish the accuracy of an eight-polar tactile-electrode impedance method in the assessment of total body water (TBW). Design: Transversal study. Setting: University department. Subjects: Fifty healthy subjects (25 men and 25 women) with a mean (s.d.) age of 40 (12)y. Methods: TBW measured by deuterium oxide dilution; resistance (R) of arms, trunk and legs measured at frequencies of 5, 50, 250 and 500 kHz with an eight-polar tactile-electrode impedance-meter (InBody 3.0, Biospace, Seoul, Korea). Results: An algorithm for the prediction of TBW from the whole-body resistance index at 500 kHz (height(2)/R-500 where R is the sum of the segmental resistances of arms, trunk and legs) was developed in a randomly chosen subsample of 35 subjects. This algorithm had an adjusted coefficient of determination (r(adj)(2)) of 0.81 (P < 0.0001) and a root mean square error (RMSE) of 3.61 (9%). Cross-validation of the predictive algorithm in the remaining 15 subjects gave an r(adj)(2) of 0.87 (P < 0.0001) and an RMSE of 3.01 (8%). The precision of eight-polar BIA, determined by measuring R three times a day for five consecutive days in a fasting subject, was < 2.8% for all segments and frequencies. Conclusion: Eight-polar BIA is a precise method that offers accurate estimates of TBW in healthy subjects. This promising method should undergo further studies of precision and its accuracy in assessing extracellular water and appenclicular body composition should be determined. Sponsorship: Modena and Reggio Emilia University.
引用
收藏
页码:1143 / 1148
页数:6
相关论文
共 17 条
[1]
ESTIMATION OF BODY-COMPOSITION FROM BIOELECTRIC IMPEDANCE OF BODY SEGMENTS [J].
BAUMGARTNER, RN ;
CHUMLEA, WC ;
ROCHE, AF .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1989, 50 (02) :221-226
[2]
The prediction of total body water and extracellular water from bioelectric impedance in obese children [J].
Bedogni, G ;
Bollea, MR ;
Severi, S ;
Trunfio, O ;
Manzieri, AM ;
Battistini, N .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 1997, 51 (03) :129-133
[3]
Bedogni G, 1996, EUR J CLIN NUTR, V50, P335
[4]
Deurenberg P, 1994, AGE NUTR, V5, P142
[5]
Glantz Stanton A., 2001, Primer of Applied Regression Analysis of Variance 3e
[6]
Guo Shumei S., 1996, American Journal of Clinical Nutrition, V64, p428S, DOI 10.1093/ajcn/64.3.428S
[7]
HEITMANN BL, 1994, EUR J CLIN NUTR, V48, P228
[8]
ANTHROPOMETRIC MEASUREMENT OF MUSCLE MASS - REVISED EQUATIONS FOR CALCULATING BONE-FREE ARM MUSCLE AREA [J].
HEYMSFIELD, SB ;
MCMANUS, C ;
SMITH, J ;
STEVENS, V ;
NIXON, DW .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1982, 36 (04) :680-690
[9]
KUSHNER RF, 1992, J AM COLL NUTR, V11, P199
[10]
IS THE IMPEDANCE INDEX (HT2/R) SIGNIFICANT IN PREDICTING TOTAL-BODY WATER [J].
KUSHNER, RF ;
SCHOELLER, DA ;
FJELD, CR ;
DANFORD, L .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1992, 56 (05) :835-839