Improved procedural results of coronary angioplasty with intravascular ultrasound-guided balloon sizing - The CLOUT pilot trial

被引:121
作者
Stone, GW
Hodgson, JM
StGoar, FG
Frey, A
Mudra, H
Sheehan, H
Linnemeier, TJ
机构
[1] HERZ ZENTRUM, BAD KROZINGEN, GERMANY
[2] UNIV MUNICH, KLINIKUM INNENSTADT, D-8000 MUNICH, GERMANY
[3] ST VINCENTS HOSP, INDIANAPOLIS, IN USA
[4] CASE WESTERN RESERVE UNIV, CORE ANGIOG & ULTRASOUND LAB, CLEVELAND, OH 44106 USA
关键词
angioplasty; remodeling; ultrasonics; imaging; coronary disease;
D O I
10.1161/01.CIR.95.8.2044
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Indiscriminate use of balloons larger than the angiographic reference segment lumen results in high rates of ischemic complications after percutaneous transluminal coronary angioplasty (PTCA). We hypothesized that angiographically unsuspected atheromatous remodeling with vessel expansion (the Glagov phenomenon) at and adjacent to PTCA target lesions would safely accommodate oversized balloons in selected patients undergoing PTCA with intravascular ultrasound (IVUS) guidance. Methods and Results After angiographically guided PTCA of 104 lesions in 102 patients, IVUS was performed, and if atheromatous remodeling was present, PTCA was repeated with larger balloons sized halfway between the lumen and external elastic membrane. Plaque occupied a mean of 51+/-15% of the angiographically ''normal'' reference segments. Further balloon upsizing by 0.25 to 1.25 mm was therefore performed in 76 lesions (73%), increasing the nominal balloon-to-artery ratio from 1.12+/-0.15 after standard PTCA to 1.30+/-0.11 after IVUS-guided PTCA (P<.0001). As a result, the angiographic minimal luminal diameter further increased from 1.95+/-0.49 to 2.21+/-0.47 mm, the % diameter stenosis fell from 28+/-15% to 18+/-14%, and the IVUS lumen area rose from 3.16+/-1.04 to 4.52+/-1.14 mm(2) (all P<.0001). The incidence of angiographic dissection was not increased after IVUS-guided balloon upsizing (37% versus 40%, P=.67), and major complications occurred in only 2 patients (1.9%). Conclusions The demonstration by IVUS of atheromatous remodeling permits the safe use of balloons traditionally considered oversized, resulting in significantly improved luminal dimensions without increased rates of dissection or ischemic complications.
引用
收藏
页码:2044 / 2052
页数:9
相关论文
共 29 条
[1]  
BALCON R, 1992, CIRCULATION, V86, P100
[2]   RESTENOSIS AFTER CORONARY ANGIOPLASTY - THE PARADOX OF INCREASED LUMEN DIAMETER AND RESTENOSIS [J].
BEATT, KJ ;
SERRUYS, PW ;
LUIJTEN, HE ;
RENSING, BJ ;
SURYAPRANATA, H ;
DEFEYTER, P ;
VANDENBRAND, M ;
LAARMAN, GJ ;
ROELANDT, J ;
VANES, GA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (02) :258-266
[3]   INTRACORONARY ULTRASOUND EVALUATION OF INTERVENTIONAL TECHNOLOGIES [J].
DAVIDSON, CJ ;
SHEIKH, KH ;
KISSLO, KB ;
PHILLIPS, HR ;
PETER, RH ;
BEHAR, VS ;
KONG, Y ;
KRUCOFF, M ;
OHMAN, EM ;
TCHENG, JE ;
STACK, RS .
AMERICAN JOURNAL OF CARDIOLOGY, 1991, 68 (13) :1305-1309
[4]  
DHAWALE PJ, 1994, CATHET DARDIOVASC DI, V33, P267
[5]  
DIMARIO C, 1995, BRIT HEART J, V73, P26
[6]   CORONARY MORPHOLOGICAL AND CLINICAL DETERMINANTS OF PROCEDURAL OUTCOME WITH ANGIOPLASTY FOR MULTIVESSEL CORONARY-DISEASE - IMPLICATIONS FOR PATIENT SELECTION [J].
ELLIS, SG ;
VANDORMAEL, MG ;
COWLEY, MJ ;
DISCIASCIO, G ;
DELIGONUL, U ;
TOPOL, EJ ;
BULLE, TM .
CIRCULATION, 1990, 82 (04) :1193-1202
[7]   A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE [J].
FISCHMAN, DL ;
LEON, MB ;
BAIM, DS ;
SCHATZ, RA ;
SAVAGE, MP ;
PENN, I ;
DETRE, K ;
VELTRI, L ;
RICCI, D ;
NOBUYOSHI, M ;
CLEMAN, M ;
HEUSER, R ;
ALMOND, D ;
TEIRSTEIN, PS ;
FISH, RD ;
COLOMBO, A ;
BRINKER, J ;
MOSES, J ;
SHAKNOVICH, A ;
HIRSHFELD, J ;
BAILEY, S ;
ELLIS, S ;
RAKE, R ;
GOLDBERG, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) :496-501
[8]   DIFFERENCES IN RESTENOSIS PROPENSITY OF DEVICES FOR TRANSLUMINAL CORONARY INTERVENTION - A QUANTITATIVE ANGIOGRAPHIC COMPARISON OF BALLOON ANGIOPLASTY, DIRECTIONAL ATHERECTOMY, STENT IMPLANTATION AND EXCIMER-LASER ANGIOPLASTY [J].
FOLEY, DP ;
MELKERT, R ;
UMANS, VA ;
DEJAEGERE, PP ;
STRIKWERDA, S ;
DEFEYTER, PJ ;
SERRUYS, PW ;
MOREL, MA ;
VANSWIJNDREGT, EM .
EUROPEAN HEART JOURNAL, 1995, 16 (10) :1331-1346
[9]   COMPENSATORY ENLARGEMENT OF HUMAN ATHEROSCLEROTIC CORONARY-ARTERIES [J].
GLAGOV, S ;
WEISENBERG, E ;
ZARINS, CK ;
STANKUNAVICIUS, R ;
KOLETTIS, GJ .
NEW ENGLAND JOURNAL OF MEDICINE, 1987, 316 (22) :1371-1375
[10]   TRANS-LUMINAL DILATATION OF CORONARY-ARTERY STENOSIS [J].
GRUNTZIG, A .
LANCET, 1978, 1 (8058) :263-263