A remote flow cell for UV absorbance detection with capillary HPLC based on a single strand fiber optic

被引:12
作者
Wang, HL [1 ]
Yi, EC [1 ]
Ibarra, CA [1 ]
Hackett, M [1 ]
机构
[1] Univ Washington, Dept Med Chem, Seattle, WA 98195 USA
关键词
D O I
10.1039/b002349n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A remote flow cell based on a single strand of fused-silica fiber optic was built for UV absorbance detection with a packed capillary HPLC system, using commercially available pumps, detection electronics (Shimadzu) and fittings. This 'off-column' flow cell design is applicable to both pressure and electro-osmotically driven systems. The goals were to minimize the linearity and light leakage problems that often limit the performance of UV absorbance detection with capillary chromatography. A linear dynamic range of 10(3) (reserpine, lambda = 220 nm), and a concentration detection limit of 5.1 x 10(-8) mol l(-1) were observed. Baseline noise was measured at 3.5 x 10(-5) absorbance units (AU), with a standard deviation of 1.7 x 10(-5) AU. The illuminated volume of approximately 3 nl was optimized for capillaries with inner diameters in the range 50-100 mu m, and flow rates from 100 nl min(-1) to 1 mu l min(-1). These modifications of readily available instrumentation have allowed the construction of a practical system for fractionating complex mixtures of peptides in small amounts, prior to mass spectrometry or additional wet chemistry steps.
引用
收藏
页码:1061 / 1064
页数:4
相关论文
共 19 条