Current-Driven Ion Fluxes of Polymeric Membrane Ion-Selective Electrode for Potentiometric Biosensing

被引:64
作者
Ding, Jiawang [1 ,2 ,3 ]
Qin, Wei [1 ]
机构
[1] Chinese Acad Sci, Key Lab Coastal Environm Proc, Yantai Inst Coastal Zone Res Sustainable Dev, Yantai 264003, Peoples R China
[2] Chinese Acad Sci, S China Sea Inst Oceanol, Guangzhou 510301, Guangdong, Peoples R China
[3] Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China
关键词
UNIFORM CONCENTRATION PROFILES; CURRENT PULSE METHOD; DETECTION LIMITS; SENSORS; POLARIZATION; COEFFICIENTS; TITRATIONS;
D O I
10.1021/ja906723h
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A polymeric membrane ion-selective electrode (ISE) for rapid potentiometric biosensing has been developed. Butyrylcholinesterase, as a target enzyme, is immobilized on the surface of the polymeric membrane. A macro-command-controlled procedure for switching between the potentiostatic and galvanostatic steps has been designed to provide a current-driven release of the butyrylcholine substrate for in situ biosensing of the enzyme and its inhibitors. This system has the flexibility of trapping substrate ions in the ISE inner filling solution and provides rapid, continuous, and reproducible measurements of enzymes and other bioanalytes involved in enzymatic systems.
引用
收藏
页码:14640 / +
页数:4
相关论文
共 20 条
[1]   Modern Potentiometry [J].
Bakker, Eric ;
Pretsch, Ernoe .
ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2007, 46 (30) :5660-5668
[2]   Selective coulometric release of ions from ion selective polymeric membranes for calibration-free titrations [J].
Bhakthavatsalam, Vishnupriya ;
Shvarev, Alexey ;
Bakker, Eric .
ANALYST, 2006, 131 (08) :895-900
[3]   Potentiometric sensing of butyrylcholinesterase based on in situ generation and detection of substrates [J].
Ding, Jiawang ;
Qin, Wei .
CHEMICAL COMMUNICATIONS, 2009, (08) :971-973
[4]   Solid-cantact electrochemical polyion sensors for monitoring peptidase activities [J].
Fordyce, Katherine ;
Shvarev, Alexey .
ANALYTICAL CHEMISTRY, 2008, 80 (03) :827-833
[5]   Direct sensing of total acidity by chronopotentiometric flash titrations at polymer membrane ion-selective electrodes [J].
Gemene, Kebede L. ;
Bakker, Eric .
ANALYTICAL CHEMISTRY, 2008, 80 (10) :3743-3750
[6]   Limitations of Current Polarization for Lowering the Detection Limit of Potentiometric Polymeric Membrane Sensors [J].
Hoefler, Lajos ;
Bedlechowicz, Iwona ;
Vigassy, Tamas ;
Gyurcsanyi, Robert E. ;
Bakker, Eric ;
Pretsch, Ernoe .
ANALYTICAL CHEMISTRY, 2009, 81 (09) :3592-3599
[7]   Determination and identification of malathion, ethion and dichlorovos using ion mobility spectrometry [J].
Jafari, Mohammad T. .
TALANTA, 2006, 69 (05) :1054-1058
[8]  
Lindner E, 1999, ELECTROANAL, V11, P695, DOI 10.1002/(SICI)1521-4109(199907)11:10/11<695::AID-ELAN695>3.3.CO
[9]  
2-7
[10]   Pulstrodes: Triple pulse control of potentiometric sensors [J].
Makarychev-Mikhailov, S ;
Shvarev, A ;
Bakker, E .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2004, 126 (34) :10548-10549