An intralaboratory quality control program for quantitation of CD34+ cells by flow cytometry

被引:16
作者
Farley, TJ
Rooney, W
Kuhns, E
Ahmed, T
Preti, RA
机构
[1] NEW YORK BLOOD CTR, CLIN SERV, VALHALLA, NY 10595 USA
[2] NEW YORK MED COLL, VALHALLA, NY 10595 USA
来源
JOURNAL OF HEMATOTHERAPY | 1997年 / 6卷 / 04期
关键词
D O I
10.1089/scd.1.1997.6.303
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study was undertaken to develop a quality control protocol to monitor instrument, operator, and CD34 assay performance, A dual level control system was established by cryopreserving aliquots of cells from peripheral blood progenitor cell (PBPC) collections exhibiting different percentages of CD34+ cells. Twenty-five samples from each control specimen were assayed to establish a control range (mean +/- 2 SD), Levey-Jennings graphs were prepared for each control specimen to plot multiple measurements of CD34%, No significant differences were observed between fresh or cryopreserved PBPC aliquots in terms of light scatter properties or CD34 antigen density within the gated cell population, Cryopreserved PBPC samples are ideal for serving as a positive methodology control for daily CD34 analysis, Furthermore, such a system can help identify problems with assay reagents, sample preparation technique, or incorrect data analysis.
引用
收藏
页码:303 / 308
页数:6
相关论文
共 18 条
  • [1] BERENSON RJ, 1991, BLOOD, V77, P1717
  • [2] Brecher M E, 1996, J Hematother, V5, P227, DOI 10.1089/scd.1.1996.5.227
  • [3] POSITIVELY SELECTED AUTOLOGOUS BLOOD CD34+ CELLS AND UNSEPARATED PERIPHERAL-BLOOD PROGENITOR CELLS MEDIATE IDENTICAL HEMATOPOIETIC ENGRAFTMENT AFTER HIGH-DOSE VP16, IFOSFAMIDE, CARBOPLATIN, AND EPIRUBICIN
    BRUGGER, W
    HENSCHLER, R
    HEIMFELD, S
    BERENSON, RJ
    MERTELSMANN, R
    KANZ, L
    [J]. BLOOD, 1994, 84 (05) : 1421 - 1426
  • [4] *CDC, 1992, MMWR, V41
  • [5] CIVIN CI, 1984, J IMMUNOL, V133, P157
  • [6] Civin CI, 1989, LEUKOCYTE TYPING, P818
  • [7] HOY T, 1994, CLIN FLOW CYTOMETRY, P193
  • [8] Johnsen H E, 1996, J Hematother, V5, P237, DOI 10.1089/scd.1.1996.5.237
  • [9] THE USE OF CONTROL CHARTS IN THE CLINICAL LABORATORY
    LEVEY, S
    JENNINGS, ER
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1950, 20 (11) : 1059 - 1066
  • [10] Lowdell MW, 1996, BONE MARROW TRANSPL, V17, P849