Determination of iodine via the spectrum of barium mono-iodide using high-resolution continuum source molecular absorption spectrometry in a graphite furnace

被引:40
作者
Huang, Mao Dong [1 ]
Becker-Ross, Helmut [1 ]
Florek, Stefan [1 ]
Okruss, Michael [1 ]
Welz, Bernhard [2 ]
Mores, Silvane [2 ]
机构
[1] ISAS Inst Analyt Sci, Dept Interface Spect, D-12489 Berlin, Germany
[2] Univ Fed Santa Catarina, Dept Quim, BR-88040900 Florianopolis, SC, Brazil
关键词
High-resolution continuum source spectrometry; Molecular absorption spectrometry; Graphite furnace; BaI molecular absorption; Iodine determination; AIR-ACETYLENE FLAME; ELECTROTHERMAL VOLATILIZATION; EMISSION SPECTRUM; MASS-SPECTROMETRY; BROMIDE; PLASMA; INDIUM; MONOIODIDE; SALTS;
D O I
10.1016/j.sab.2009.06.010
中图分类号
O433 [光谱学];
学科分类号
070207 [光学];
摘要
Molecular absorption spectra of the diatomic molecules All, GaI, InI, TII, MgI, CaI, SrI and BaI, generated in a graphite furnace, were studied using a high-resolution echelle spectrometer with the aim of finding a simple, reliable and sensitive analytical method for the determination of iodine. Among them, the barium mono-iodide (BaI) was found to have the strongest absorption bands around 538 rim and 560 nm, each of then) consisting of a series of well-resolved rotational lines with half-widths of about 40-50 pm. The strongest BaI line, the hand head at 538.308 nm has been evaluated systematically for its analytical use for the determination of iodine. High concentrations of hydrochloric acid (or chloride), hydrofluoric acid (or fluoride). iron. potassium and sodium resulted in significant reduction of the BaI molecular absorption. Apart from this, no other serious spectral or non-spectral interference has been observed. Different chemical forms of iodine, such as iodide, iodate and organically bound iodine produced identical BaI absorption sensitivity. The detection limit for iodine was 600 pg, and the calibration curve was linear up to 250 ng iodine. Two real samples with different chemical forms of iodine were analyzed using the proposed method. One sample was an iodide pill with a specified iodide content of 200 mg, the other one was a thyroid hormone pill with a specified content of 63.5 mg. The results were in good or satisfactory agreement with those of independent methods, the potentiometric titration and the inductively coupled plasma time-of-flight mass spectrometry (ICP-ToF-MS); the deviations were 2% and 8% for the iodide and the thyroid hormone sample, respectively. The relative standard deviation of the analytical results (n = 3) was below 2%. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:697 / 701
页数:5
相关论文
共 41 条
[1]
DETERMINATION OF IODINE AND BROMINE IN PLASMA AND URINE BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
MAURAS, Y ;
DOUGE, C ;
JAUNAULT, L ;
DELAPORTE, T ;
BEAUGRAND, C .
ANALYST, 1990, 115 (06) :813-815
[2]
[Anonymous], 1997, CRC HDB CHEM PHYS
[3]
[Anonymous], 1994, Schistosomes, liver flukes and Helicobacter pylori, V61, P1
[4]
ASHRAFUNNISA DVK, 1973, J PHYS B ATOM MOL PH, V5, P1503
[5]
PRODUCTION OF COPPER(I) HALIDE EMISSION-SPECTRA BY MECA [J].
BELCHER, R ;
BOGDANSK.SL ;
GHONAIM, SA ;
TOWNSHEN.A .
NATURE, 1974, 248 (5446) :326-327
[6]
Indirect atomic absorption spectrometry (IAAS) as a tool for the determination of iodide in infant formulas by precipitation of AgI and redissolution with cyanide [J].
Bermejo-Barrera, P ;
Fernández-Sánchez, LM ;
Aboal-Somoza, M ;
Anllo-Sendín, RM ;
Bermejo-Barrera, A .
MICROCHEMICAL JOURNAL, 2001, 69 (03) :205-211
[7]
MOLECULAR-EMISSION SPECTROSCOPY IN COOL FLAMES .4. DETERMINATION OF CHLORIDE, BROMIDE AND IODIDE BY THERMAL-EMISSION SPECTROSCOPY IN PRESENCE OF INDIUM SALTS [J].
DAGNALL, RM ;
THOMPSON, KC ;
WEST, TS .
ANALYST, 1969, 94 (1121) :643-&
[8]
MOLECULAR ABSORPTION SPECTROMETRY WITH ELECTROTHERMAL VOLATILIZATION IN A GRAPHITE TUBE .5. DETERMINATION OF IODIDE TRACES BY MOLECULAR ABSORPTION OF TLI [J].
DITTRICH, K ;
SCHNEIDER, S .
ANALYTICA CHIMICA ACTA, 1980, 115 (MAR) :201-209
[10]
The determination of iodine species in environmental and biological samples (Technical report) [J].
Edmonds, JS ;
Morita, M .