The new Frenchay artificial cervical joint - Results from a two-year pilot study

被引:176
作者
Wigfield, CC
Gill, SS
Nelson, RJ
Metcalf, NH
Robertson, JT
机构
[1] Frenchay Hosp, Spinal Res & Dev Unit, Acad Ctr, Dept Neurosurg, Bristol BS16 1LE, Avon, England
[2] Medtron Sofamor Danek, Clin & Regulatory Affairs, Memphis, TN USA
[3] Univ Tennessee, Dept Neurosurg, Memphis, TN USA
关键词
artificial; cervical; disc; intervertebral; joint; motion;
D O I
10.1097/00007632-200211150-00006
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. A clinical prospective observational cohort study of 15 patients undergoing cervical intervertebral disc replacement. Objective. To assess the safety, clinical stability, and capacity of a newly designed cervical intervertebral disc replacement for preserving motion in the cervical spine of patients with degenerative disc disease. Summary of Background Data. Anterior cervical discectomy and interbody fusion results in loss of motion segments and there is evidence of accelerated degenerative changes occurring at adjacent levels. Intervertebral disc replacement may be a valid alternative to fusion. There is a need for cervical disc replacement to be evaluated in prospective studies before it can be adopted as an acceptable surgical technique. Methods. The study enrolled 15 patients with cervical radiculopathy or myelopathy and radiologically confirmed cervical disc herniation or posterior vertebral body osteophytes. Eligibility for the study required that patients have either a previous adjacent-level surgical or congenital spinal fusion or radiological evidence of adjacent-level degenerative disc disease. After decompressive surgery via an anterior approach, all 15 patients received the artificial joint. Follow-up visits were at set intervals and included clinical examination, radiological assessment, and evaluation by patient-completed questionnaires. Results. In all cases, the artificial joint maintained motion at the operative levels while reestablishing intervertebral height. The procedure was considered safe for experienced spine surgeons to perform, and the device was stable, with no dislocation or components or backing out of screws. Two screws broke, but without any consequence. Improvements in assessment scores were noted. Conclusions. Cervical intervertebral motion can be maintained with the new device, which is clinically stable. Meticulous attention must be paid to the surgical technique to maximize the chances of a good result. The pilot study was successful, although it has yet to be determined what conditions will benefit most from this technology.
引用
收藏
页码:2446 / 2452
页数:7
相关论文
共 26 条
[1]
POSTFUSION INSTABILITY AT THE ADJACENT SEGMENTS AFTER RIGID PEDICLE SCREW FIXATION FOR DEGENERATIVE LUMBAR SPINAL-DISORDERS [J].
AOTA, Y ;
KUMANO, K ;
HIRABAYASHI, S .
JOURNAL OF SPINAL DISORDERS, 1995, 8 (06) :464-473
[2]
The artificial disc: Theory, design and materials [J].
Bao, QB ;
McCullen, GM ;
Higham, PA ;
Dumbleton, JH ;
Yuan, HA .
BIOMATERIALS, 1996, 17 (12) :1157-1167
[3]
CHERUBINO P, 1990, Italian Journal of Orthopaedics and Traumatology, V16, P533
[4]
THE ANTERIOR APPROACH FOR REMOVAL OF RUPTURED CERVICAL DISKS [J].
CLOWARD, RB .
JOURNAL OF NEUROSURGERY, 1958, 15 (06) :602-617
[5]
Surgical experience with an implanted artificial cervical joint [J].
Cummins, BH ;
Robertson, JT ;
Gill, SS .
JOURNAL OF NEUROSURGERY, 1998, 88 (06) :943-948
[6]
DIANGELO, 1998, 6 ANN CERV SPIN RES
[7]
INVIVO FLEXION EXTENSION OF THE NORMAL CERVICAL-SPINE [J].
DVORAK, J ;
PANJABI, MM ;
NOVOTNY, JE ;
ANTINNES, JA .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1991, 9 (06) :828-834
[8]
ANTERIOR CERVICAL FUSION FOR DEGENERATED OR PROTRUDED DISKS - A REVIEW OF 146 PATIENTS [J].
GORE, DR ;
SEPIC, SB .
SPINE, 1984, 9 (07) :667-671
[9]
ROENTGENOGRAPHIC FINDINGS OF THE CERVICAL-SPINE IN ASYMPTOMATIC PEOPLE [J].
GORE, DR ;
SEPIC, SB ;
GARDNER, GM .
SPINE, 1986, 11 (06) :521-524
[10]
GRUNDY P, 1998, LONG TERM OUTCOME AN