Inter-laboratory validation of the fluorescent protein phosphatase inhibition assay to determine diarrhetic shellfish-toxins:: intercomparison with liquid chromatography and mouse bioassay

被引:18
作者
González, JC
Leira, F
Fontal, OI
Vieytes, MR
Arévalo, FF
Vieites, JM
Bermúdez-Puente, M
Muñiz, S
Salgado, C
Yasumoto, T
Botana, LM
机构
[1] Fac Vet Sci, Dept Pharmacol, Lugo 27002, Spain
[2] Fac Vet Sci, Dept Physiol, Lugo 27002, Spain
[3] ANFACO CECOPESCA, Vigo 36310, Spain
[4] Ctr Control Calidade Medio Marino, Pontevedra 36611, Spain
[5] Japan Food Labs, Tama Lab, Tokyo 2060025, Japan
关键词
D O I
10.1016/S0003-2670(02)00597-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Toxic episodes of diarrhetic shellfish toxins (DSP) in shellfish harvesting areas have serious economic and public health implications, where fluorescent protein phosphatase inhibition assay (FPPIA) may be a highly useful tool for monitoring purposes. This paper presents results from the first inter-laboratory study to validate the assay. Three laboratories participated in the design and development of the inter-laboratory work. Standard solutions and spiked samples of the main toxin, okadaic acid, were used at the beginning of the validation exercise to avoid cross-inhibition of other toxins that would otherwise deteriorate the quantitative significance of the data. HPLC with fluorimetric detection of okadaic acid was also submitted to inter-laboratory validation to be subsequently used as a quantitative reference method. FPPIA results from spiked samples were free of systematic bias in any laboratory and determinations repeated over 3 days showed that the classic "repeatability" was the main within-laboratory source of variability (15-26% R.S.D. depending on the sample). After the inter-laboratory validation of both HPLC and FPPIA methods, 83 samples of mussel hepatopancreas collected during a toxic DSP episode were analyzed over 9 weeks. Toxic levels determined with FPPIA were in line with mouse bioassay results, highlighting the lack of false negative results of the FPPIA test: 98.7% of samples whose concentration of okadaic acid equivalents was over 0.8 mug/g hep., provided positive bioassay results within 24h of observation time. The reliability and the quantitativeness of the FPPIA method in naturally contaminated samples was demonstrated by intercomparison with mouse bioassay and HPLC. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:233 / 246
页数:14
相关论文
共 18 条
[1]  
*ASS OFF AN CHEM I, 1995, J ASSOC OFF ANA CHEM, V78, pA143
[2]   PROCESSING DATA FOR OUTLIERS [J].
DIXON, WJ .
BIOMETRICS, 1953, 9 (01) :74-89
[3]  
FONTAL OI, 1997, P 8 INT C HARMF ALG, P299
[4]   Development and validation of a high-performance liquid chromatographic method using fluorimetric detection for the determination of the diarrhetic shellfish poisoning toxin okadaic acid without chlorinated solvents [J].
González, JC ;
Leira, F ;
Vieytes, MR ;
Vieites, JM ;
Botana, AM ;
Botana, LM .
JOURNAL OF CHROMATOGRAPHY A, 2000, 876 (1-2) :117-125
[5]  
GREEN JM, 1996, ANAL CHEM, V68, P305
[6]   FLUOROMETRIC-DETERMINATION OF DIARRHETIC SHELLFISH TOXINS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
LEE, JS ;
YANAGI, T ;
KENMA, R ;
YASUMOTO, T .
AGRICULTURAL AND BIOLOGICAL CHEMISTRY, 1987, 51 (03) :877-881
[7]   QUANTIFICATION OF DIARRHETIC SHELLFISH TOXINS AND IDENTIFICATION OF NOVEL PROTEIN PHOSPHATASE INHIBITORS IN MARINE-PHYTOPLANKTON AND MUSSELS [J].
LUU, HA ;
CHEN, DZX ;
MAGOON, J ;
WORMS, J ;
SMITH, J ;
HOLMES, CFB .
TOXICON, 1993, 31 (01) :75-83
[8]   CYANOBACTERIAL MICROCYSTIN-LR IS A POTENT AND SPECIFIC INHIBITOR OF PROTEIN PHOSPHATASE-1 AND PHOSPHATASE-2A FROM BOTH MAMMALS AND HIGHER-PLANTS [J].
MACKINTOSH, C ;
BEATTIE, KA ;
KLUMPP, S ;
COHEN, P ;
CODD, GA .
FEBS LETTERS, 1990, 264 (02) :187-192
[9]   Evaluation of the fluorometric protein phosphatase inhibition assay in the determination of okadaic acid in mussels [J].
Mountfort, DO ;
Kennedy, G ;
Garthwaite, I ;
Quilliam, M ;
Truman, P ;
Hannah, DJ .
TOXICON, 1999, 37 (06) :909-922
[10]  
*NAT I STAND TECHN, 1994, 80 NIST