Is the placement of a pulmonary artery catheter still justified solely for the measurement of cardiac output?

被引:54
作者
Sakka, SG
Reinhart, K
Wegscheider, K
Meier-Hellmann, A
机构
[1] Univ Jena, Dept Anesthesiol & Crit Care Med, D-07740 Jena, Germany
[2] Univ Hamburg, Dept Stat & Econometry, Hamburg, Germany
关键词
cardiac output; critically ill patients; pulmonary artery thermodilution; transpulmonary thermodilution; Fick principle;
D O I
10.1016/S1053-0770(00)90002-8
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective: The authors compared four clinical techniques of measuring cardiac output (CO) in critically ill patients: pulmonary artery thermodilution (CO[PA]), transpulmonary aortic thermodilution (CO[AORTA]), Fick principle-derived (CO[FICK]), and continuous pulmonary artery (CCO) measurements. Design: Prospective clinical study. Setting: Surgical intensive care unit of a university hospital. Participants: Twelve adult patients suffering from sepsis or septic shock. Interventions: All patients were deeply sedated and mechanically ventilated in a pressure-controlled mode. Each patient received a 7.5F five-lumen pulmonary artery catheter for the continuous measurement of cardiac output and a 4F aortic catheter with an integrated thermistor. The thermistors of the two different catheters were connected to one computer system (COLD-Z021, Pulsion Medical Systems, Munich, Germany). Whole-body oxygen consumption was measured by indirect calorimetry using a metabolic cart (Deltatrac, Datex-Engstroem, Helsinki, Finland) over a 5-minute period, at the end of which arterial and mixed venous blood gases were taken and measured by co-oximetry. During each measuring period, three bolus CO measurements were performed. A total number of 51 CO measurements was analyzed. Results: Linear regression analysis revealed the highest correlation between CO(AORTA) and CO(PA) (r = 0.98), whereas agreement between these two techniques and CCO was lower (r = 0.92 and r = 0.93). All three techniques correlated comparably with CO(FICK) (r = 0.85, r = 0.83, and r = 0.83). Conclusion: The correlations among the four CO techniques were high and similar, with CO(PA) and CO(AORTA) techniques showing the highest agreement. Because CO with similar accuracy can be obtained from transpulmonary aortic thermodilution in a less-invasive manner, it appears that the placement of a pulmonary artery catheter solely for the measurement of CO is no longer justified, unless continuous CO measurements are needed. Copyright (C) 2000 by W.B. Saunders Company.
引用
收藏
页码:119 / 124
页数:6
相关论文
共 31 条
[1]   CONSERVATION OF THERMAL INDICATOR IN LUNG CIRCULATION [J].
ARFORS, K ;
MALMBERG, P ;
PAVEK, K .
CARDIOVASCULAR RESEARCH, 1971, 5 (04) :530-+
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   CARDIAC-OUTPUT MEASUREMENT USING FEMORAL-ARTERY THERMODILUTION IN PATIENTS [J].
BOCK, JC ;
BARKER, BC ;
MACKERSIE, RC ;
TRANBAUGH, RF ;
LEWIS, FR .
JOURNAL OF CRITICAL CARE, 1989, 4 (02) :106-111
[4]   AMERICAN-COLLEGE OF CHEST PHYSICIANS SOCIETY OF CRITICAL CARE MEDICINE CONSENSUS CONFERENCE - DEFINITIONS FOR SEPSIS AND ORGAN FAILURE AND GUIDELINES FOR THE USE OF INNOVATIVE THERAPIES IN SEPSIS [J].
BONE, RC ;
BALK, RA ;
CERRA, FB ;
DELLINGER, RP ;
FEIN, AM ;
KNAUS, WA ;
SCHEIN, RMH ;
SIBBALD, WJ ;
ABRAMS, JH ;
BERNARD, GR ;
BIONDI, JW ;
CALVIN, JE ;
DEMLING, R ;
FAHEY, PJ ;
FISHER, CJ ;
FRANKLIN, C ;
GORELICK, KJ ;
KELLEY, MA ;
MAKI, DG ;
MARSHALL, JC ;
MERRILL, WW ;
PRIBBLE, JP ;
RACKOW, EC ;
RODELL, TC ;
SHEAGREN, JN ;
SILVER, M ;
SPRUNG, CL ;
STRAUBE, RC ;
TOBIN, MJ ;
TRENHOLME, GM ;
WAGNER, DP ;
WEBB, CD ;
WHERRY, JC ;
WIEDEMANN, HP ;
WORTEL, CH .
CRITICAL CARE MEDICINE, 1992, 20 (06) :864-874
[5]   Effects of continuous negative extrathoracic pressure versus positive end-expiratory pressure in acute lung injury patients [J].
Borelli, M ;
Benini, A ;
Denkewitz, T ;
Acciaro, C ;
Foti, G ;
Pesenti, A .
CRITICAL CARE MEDICINE, 1998, 26 (06) :1025-1031
[6]   RELATIONSHIP BETWEEN CO AND TRANSIT TIMES FOR DYE AND THERMAL INDICATORS IN CENTRAL CIRCULATION [J].
CARLILE, PV ;
BECKETT, RC ;
GRAY, BA .
JOURNAL OF APPLIED PHYSIOLOGY, 1986, 60 (04) :1363-1372
[7]   The effectiveness of right heart catheterization in the initial care of critically ill patients [J].
Connors, AF ;
Speroff, T ;
Dawson, NV ;
Thomas, C ;
Harrell, FE ;
Wagner, D ;
Desbiens, N ;
Goldman, L ;
Wu, AW ;
Califf, RM ;
Fulkerson, WJ ;
Vidaillet, H ;
Broste, S ;
Bellamy, P ;
Lynn, J ;
Knaus, WA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11) :889-897
[8]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[9]   Radial artery pressure monitoring underestimates central arterial pressure during vasopressor therapy in critically ill surgical patients [J].
Dorman, T ;
Breslow, MJ ;
Lipsett, PA ;
Rosenberg, JM ;
Balser, JR ;
Almog, Y ;
Rosenfeld, BA .
CRITICAL CARE MEDICINE, 1998, 26 (10) :1646-1649
[10]   Reproducibility of double indicator dilution measurements of intrathoracic blood volume compartments, extravascular lung water, and liver function [J].
Godje, O ;
Peyerl, M ;
Seebauer, T ;
Dewald, O ;
Reichart, B .
CHEST, 1998, 113 (04) :1070-1077