Comparison of different approaches for evaluation of the detection and quantitation limits of a purity method: A case study using a capillary isoelectrofocusing method for a monoclonal antibody

被引:37
作者
Apostol, Izydor [1 ]
Miller, Karen J. [1 ]
Ratto, Joseph [1 ]
Kelner, Drew N. [1 ]
机构
[1] Amgen Inc, Analyt & Formulat Sci, Thousand Oaks, CA 91320 USA
关键词
LOD; LOQ; Purity method; cIEF separation; Monoclonal antibody;
D O I
10.1016/j.ab.2008.09.053
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Several different techniques suggested by the International Conference on Harmonization (ICH) Q2R1 guideline were used to assess the signal and concentration at the limit of detection (LOD) and limit of quantitation (LOQ) for a purity method. These approaches were exemplified with a capillary isoelectro-focusing (cIF) method, which has been developed to quantify the distribution of the charge isoforms of a monoclonal antibody. The charge isoforms are the result of incomplete posttranslational processing of C-terminal lysine residues of the heavy chain by carboxypeptidase. Results showed no significant discrepancy between LOD/LOQ obtained by the different techniques. Validation experiments corroborated the calculated LOQ. The results indicate that any single technique can provide meaningful values for the LOD and LOQ. Finally, important points to consider when applying these techniques to purity methods are discussed. (C) 2008 Elsevier Inc. All rights reserved.
引用
收藏
页码:101 / 106
页数:6
相关论文
共 12 条
[1]   A heuristic derivation of the Horwitz curve [J].
Albert, R ;
Horwitz, W .
ANALYTICAL CHEMISTRY, 1997, 69 (04) :789-790
[2]  
[Anonymous], 2005, Quality Risk Management Q9, P1
[3]  
Apostol I, 2008, BIOPROCESS INT, V6, P12
[4]   PROCESSING OF C-TERMINAL LYSINE AND ARGININE RESIDUES OF PROTEINS ISOLATED FROM MAMMALIAN-CELL CULTURE [J].
HARRIS, RJ .
JOURNAL OF CHROMATOGRAPHY A, 1995, 705 (01) :129-134
[5]  
HORWITZ W, 1982, ANAL CHEM, V54, pA67, DOI 10.1021/ac00238a002
[6]  
Miller JN., 2000, STAT CHEMOMETRICS AN, V4th
[7]  
Motulsky H., 1995, Intuitive biostatistics
[8]   Determination of the origin of charge heterogeneity in a murine monoclonal antibody [J].
Perkins, M ;
Theiler, R ;
Lunte, S ;
Jeschke, M .
PHARMACEUTICAL RESEARCH, 2000, 17 (09) :1110-1117
[9]  
Rubinson K.A., 2000, Contemporary Instrumental Analysis, V1st
[10]   Downstream processing of monoclonal antibodies - Application of platform approaches [J].
Shukla, Abhinav A. ;
Hubbard, Brian ;
Tressel, Tim ;
Guhan, Sam ;
Low, Duncan .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 848 (01) :28-39