Failure analysis of total knee replacement. Basics and methodological aspects of the damage analysis

被引:20
作者
Bader, R.
Mittelmeier, W.
Steinhauser, E.
机构
[1] Univ Rostock, Orthopad Klin & Poliklin, D-18057 Rostock, Germany
[2] Tech Univ Munich, Klin Orthopad & Unfallchirurg, D-8000 Munich, Germany
来源
ORTHOPADE | 2006年 / 35卷 / 09期
关键词
total knee endoprosthesis; failure; revision; damage analysis; methods;
D O I
10.1007/s00132-006-0976-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Possible causes for failure of total knee endoprostheses represent wear, malpositioning, maldimensioning and inadequate design of the implant components, manufacturing defects, material fatigue, corrosion, overloading, infection, and allergy against implant materials. There is a broad spectrum of methodical approaches for the analysis of failure cases. Substantial information for the damage analysis is provided by clinical and intraoperative findings, photo documentation, radiographic course as well as all-solid, physical and histological investigations. Principal purposes of damage analysis are the avoidance of further damage events and the gain of information for improvement of implant design and material as well as the optimisation of the biocompatibility of implants and wear products. Both a detection system of incidents and implant failures as well as a complete data collection enables early identification of system-specific, accumulated cases of implant failure.
引用
收藏
页码:896 / +
页数:6
相关论文
共 23 条
[1]
ALBERT K, 1994, BIO-MED MATER ENG, V4, P199
[2]
[Anonymous], 67 ANN M AM AC ORTH
[3]
Bargel H. J., 1987, WERKSTOFFKUNDE
[4]
CAMERON HU, 1991, CAN J SURG, V34, P625
[5]
Graf R, 2003, ORTHOPADE, V32, P453, DOI 10.1007/s00132-003-0506-2
[6]
HARRIS WH, 1995, CLIN ORTHOP RELAT R, P46
[7]
The Norwegian Arthroplasty Register -: 11 years and 73,000 arthroplasties [J].
Havelin, LI ;
Engesæter, LB ;
Espehaug, B ;
Furnes, O ;
Lie, SA ;
Vollset, SE .
ACTA ORTHOPAEDICA SCANDINAVICA, 2000, 71 (04) :337-353
[8]
Survey of incidents associated with hip and knee replacement devices. Analysis of the surveillance and registration system for medical products for the years 2000-2002 [J].
Heinz, BC ;
von Mallek, D .
ORTHOPADE, 2005, 34 (01) :47-+
[9]
Hofmann S, 2003, ORTHOPADE, V32, P469, DOI 10.1007/s00132-003-0503-5
[10]
HUISKES R, 1990, CLIN ORTHOP RELAT R, P27