Accounting for incomplete charge collection in Monte Carlo simulations of the efficiency of well-type Ge-detectors

被引:22
作者
Hernández, F [1 ]
El-Daoushy, F [1 ]
机构
[1] Uppsala Univ, Dept Phys, S-75121 Uppsala, Sweden
关键词
Monte Carlo simulations; HPGe detectors; efficiency; incomplete charge collection;
D O I
10.1016/S0168-9002(02)02080-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The influence of the process of incomplete charge collection on the performance properties of HPGe well-type detectors is studied. A simplified sub-routine, that takes into account the total collection time of charges within the Ge-crystal, has been developed and coupled to the Monte Carlo photopeak efficiency. Only the manufacturer's data of the detector's geometry and the detector's operating voltage was used in the sub-routine. The corrected photopeak efficiency simulations, as based on this new sub-routine, showed much better agreement with the experimental data, especially for energies above 40 keV. The ratio between the experimental and simulated efficiency at 1460 keV were, for example, improved from 0.73 to 0.96 after performing these corrections on our largest well-type Ge-detector. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:340 / 351
页数:12
相关论文
共 21 条
[1]   Precise modeling of a coaxial HPGe detector [J].
Ashrafi, S ;
Likar, A ;
Vidmar, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 438 (2-3) :421-428
[2]   Ge detector efficiency calculations for determination of three-dimensional radioactivity distributions [J].
Baas, HW ;
Bode, P ;
Blaauw, M .
RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) :637-638
[3]  
Briesmeister JF., 1997, ALAMOS NATL LAB REPO, pLA
[4]  
BRONSON F, 1996, WAST MAN 96 TUCS AZ
[5]  
*EG G ORTEC, 1994, CAT MOD PULS PROC EL
[6]   Gamma spectrometry of 234Th (238U) in environmental samples [J].
El-Daoushy, F ;
Hernández, F .
ANALYST, 2002, 127 (07) :981-989
[7]   WELL GE AND SEMI-PLANAR GE (HP) DETECTORS FOR LOW-LEVEL GAMMA-SPECTROMETRY [J].
ELDAOUSHY, F ;
GARCIATENORIO, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 356 (2-3) :376-384
[8]   Monte Carlo determination of full energy peak efficiency for a HPGe detector [J].
Ewa, IOB ;
Bodizs, D ;
Czifrus, S ;
Molnar, Z .
APPLIED RADIATION AND ISOTOPES, 2001, 55 (01) :103-108
[9]   MONTE-CARLO SIMULATION OF SEVERAL GAMMA-EMITTING SOURCE AND DETECTOR ARRANGEMENTS FOR DETERMINING CORRECTIONS OF SELF-ATTENUATION AND COINCIDENCE SUMMATION IN GAMMA-SPECTROMETRY [J].
HAASE, G ;
TAIT, D ;
WIECHEN, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 329 (03) :483-492
[10]   Semi-empirical method for self-absorption correction of photons with energies as low as 10 keV in environmental samples [J].
Hernández, F ;
El-Daoushy, F .