Determination of trace level bromate and perchlorate in drinking water by ion chromatography with an evaporative preconcentration technique

被引:47
作者
Liu, YJ
Mou, S
Heberling, S
机构
[1] Chinese Acad Sci, Ecoenvironm Sci Res Ctr, Beijing 100085, Peoples R China
[2] Dionex Chem Corp, Sunnyvale, CA 94086 USA
关键词
D O I
10.1016/S0021-9673(01)01507-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple sample preconcentration technique employing microwave-based evaporation for the determination of trace level bromate and perchlorate in drinking water with ion chromatography is presented. With a hydrophilic anion-exchange column and a sodium hydroxide eluent in linear gradient, bromate and perchlorate can be determined in one injection within 35 min. Prior to ion chromatographic analysis, the drinking water sample was treated with an OnGuard-Ag cartridge to remove the superfluous chloride and concentrated 20-fold using a PTFE beaker in a domestic microwave oven for 15 min. The recoveries of the anions ranged from 94.6% for NO2- to 105.2% for F-. The detection limits for bromate, perchlorate, iodate and chlorate were 0.1, 0.2, 0.1 and 0.2 mug/l, respectively. The developed method is applicable for the quantitation of bromate and perchlorate in drinking water samples. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:85 / 91
页数:7
相关论文
共 27 条
[1]   Ion chromatographic determination of bromate in drinking water by post-column reaction with fuchsin [J].
Achilli, M ;
Romele, L .
JOURNAL OF CHROMATOGRAPHY A, 1999, 847 (1-2) :271-277
[2]  
*CAL DEP HLTH SERY, 1997, PERCHL CAL DRINK WAT
[3]   Electrospray ion chromatography tandem mass spectrometry of oxyhalides at sub-ppb levels [J].
Charles, L ;
Pépin, D .
ANALYTICAL CHEMISTRY, 1998, 70 (02) :353-359
[4]   Electrospray ion chromatography tandem mass spectrometry of bromate at sub-ppb levels in water [J].
Charles, L ;
Pepin, D ;
Casetta, B .
ANALYTICAL CHEMISTRY, 1996, 68 (15) :2554-2558
[5]  
DOLLARHIDE JS, 1995, COMMUNICATION
[6]  
*EPA, 1998, 815F98002 EPA
[7]   ANALYSIS OF OXYHALIDE DISINFECTION BY-PRODUCTS AND OTHER ANIONS OF INTEREST IN DRINKING-WATER BY ION CHROMATOGRAPHY [J].
HAUTMAN, DP ;
BOLYARD, M .
JOURNAL OF CHROMATOGRAPHY, 1992, 602 (1-2) :65-74
[8]   Determination of bromate in drinking water at the low μg/L level by column switching ion chromatography [J].
Huang, YA ;
Mou, SF ;
Yan, Y .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 1999, 22 (14) :2235-2245
[9]   Determination of trace level perchlorate in drinking water and ground water by ion chromatography [J].
Jackson, PE ;
Laikhtman, M ;
Rohrer, JS .
JOURNAL OF CHROMATOGRAPHY A, 1999, 850 (1-2) :131-135
[10]   TRACE-LEVEL DETERMINATION OF BROMATE IN OZONATED DRINKING-WATER USING ION CHROMATOGRAPHY [J].
JOYCE, RJ ;
DHILLON, HS .
JOURNAL OF CHROMATOGRAPHY A, 1994, 671 (1-2) :165-171