A new method for measurement of blood pressure, heart rate, and activity in the mouse by radiotelemetry

被引:126
作者
Mills, PA
Huetteman, DA
Brockway, BP
Zwiers, LM
Gelsema, AJM
Schwartz, RS
Kramer, K
机构
[1] Data Sci Int, St Paul, MN 55126 USA
[2] IBIS Instrumentat Canada, Ottawa, ON K1Z 7Z3, Canada
[3] Mayo Clin, Rochester, MN 55905 USA
[4] Free Univ Amsterdam, Leiden Amsterdam Ctr Drug Res, Div Mol Pharmacol, Dept Pharmacochem, NL-1081 HV Amsterdam, Netherlands
关键词
conscious animals; chronic implant; frequency response; biocompatibility;
D O I
10.1152/jappl.2000.88.5.1537
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
A simple and reliable means for accurate, chronic measurement bf pulsatile blood pressure (BP) from conscious, freely moving laboratory mice was developed and validated. The newly developed device consists of a small (1.9 mi, 3.4 g), fully implantable radiotelemetry transmitter. Initial frequency response tests showed an adequate dynamic response; the average -3-dB point found in five transmitters was 145 +/- 14 (SD) Hz. BP, heart rate, and locomotor activity were recorded from 16 chronically (30-150 days) implanted mice. Mean arterial and pulse pressure, checked at regular intervals, ranged from 90-140 mmHg and from 30-50 mmHg, respectively, throughout the study. Transmitter BP measurements were validated against a Millar 1.4-Fr. transducer-tipped catheter. The mean error of the transmitters for diastolic pressures was +1.1 +/- 6.9 mmHg (n = 7). The error for systolic pressures was, on average, 2.7 +/- 3.9 mmHg larger. This new device accurately monitors BP, heart rate, and locomotor activity in conscious, untethered, freely moving mice living in their home cages for periods of at least 150 days.
引用
收藏
页码:1537 / 1544
页数:8
相关论文
共 16 条
[1]  
[Anonymous], 1991, HDB BLOOD PRESSURE M
[2]  
BAUMANS V, 1998, NOLDUS INFORMATION T
[3]   TELEMETRIC MONITORING OF CARDIOVASCULAR PARAMETERS IN CONSCIOUS SPONTANEOUSLY HYPERTENSIVE RATS [J].
BAZIL, MK ;
KRULAN, C ;
WEBB, RL .
JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1993, 22 (06) :897-905
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
BROCKWAY BP, 1993, NEW TECHNOLOGIES CON, P109
[6]  
BROCKWAY BP, 1989, Patent No. 4846191
[7]  
BUNAG RD, 1983, CLIN EXP HYPERTENS A, V5, P1659
[8]   LACK OF CORRELATION BETWEEN DIRECT AND INDIRECT MEASUREMENTS OF ARTERIAL PRESSURE IN UNANAESTHETIZED RATS [J].
BUNAG, RD ;
MCCUBBIN, JW ;
PAGE, IH .
CARDIOVASCULAR RESEARCH, 1971, 5 (01) :24-&
[9]   NONINVASIVE MEASUREMENT OF ASCENDING AORTIC BLOOD VELOCITY IN MICE [J].
HARTLEY, CJ ;
MICHAEL, LH ;
ENTMAN, ML .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 268 (01) :H499-H505
[10]   USE OF TELEMETRY TO RECORD ELECTROCARDIOGRAM AND HEART-RATE IN FREELY MOVING MICE [J].
KRAMER, K ;
VANACKER, SABE ;
VOSS, HP ;
GRIMBERGEN, JA ;
VANDERVIJGH, WJF ;
BAST, A .
JOURNAL OF PHARMACOLOGICAL AND TOXICOLOGICAL METHODS, 1993, 30 (04) :209-215