Buttressing angle of the double-plating fixation of a distal radius fracture: a finite element study

被引:41
作者
Lin, Chun-Li
Lin, Yu-Hao
Chen, Alvin Chao-Yu
机构
[1] Chang Gung Mem Hosp, Dept Orthopaed Surg, Tao Yuan, Taiwan
[2] Chang Gung Univ, Grad Inst Med Mechatron, Tao Yuan 333, Taiwan
[3] Chang Gung Univ, Dept Mech Engn, Tao Yuan 333, Taiwan
关键词
distal radius fracture; finite element method; double-plating; biomechanics; fixation plate;
D O I
10.1007/s11517-006-0082-9
中图分类号
TP39 [计算机的应用];
学科分类号
081203 [计算机应用技术]; 0835 [软件工程];
摘要
Treatment of a distal radius fracture should consider principles including stable fixation and early motion. The aim of this study was to investigate the biomechanical interactions of plate-fixation angles in the internal double-plating method coupled with various load conditions using non-linear finite element analysis (FEA). A 3D finite element distal radius fracture model with three separation angles (50, 70, and 90 degrees) between the buttressed L- and straight plates was generated based on computed tomography data. After model verification and validation, frictional (contact) elements were used to simulate the interface condition between the fixation plates and the bony surface. The stress/strain distributions and displacements at the radius end were observed under axial, bending, and torsion load conditions. The simulated results indicated that the bending and torsion increased the stress values more than the axial load. The radius and straight plate stress values decreased significantly with increasing fixation angles for all load conditions. However, the L-plate stress values increased slightly under the bending buckling effect. The displacements at the radius end and strains at the fracture healing interface decreased with increasing fixation angles for axial and torsion conditions but displayed a slight difference for the bending condition. The findings using FEA provide quantitative evidence to identify that much larger plate fixation angles could provide better mechanical strength to establish favorable stress-transmission and prevent distal fragment dislocation.
引用
收藏
页码:665 / 673
页数:9
相关论文
共 31 条
[1]
Stress analysis of interbody fusion-finite element modelling of intervertebral implant and vertebral body [J].
Adam, C ;
Pearcy, M ;
McCombe, P .
CLINICAL BIOMECHANICS, 2003, 18 (04) :265-272
[2]
Initial stability of ankle arthrodesis with three-screw fixation.: A finite element analysis [J].
Alonso-Vázquez, A ;
Lauge-Pedersen, H ;
Lidgren, L ;
Taylor, M .
CLINICAL BIOMECHANICS, 2004, 19 (07) :751-759
[3]
FORCES IN THE NORMAL AND ABNORMAL HAND [J].
AN, KN ;
CHAO, EY ;
COONEY, WP ;
LINSCHEID, RL .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1985, 3 (02) :202-211
[4]
Intramedullary nails with two lag screws [J].
Brown, CJ ;
Wang, CJ ;
Yettram, AL ;
Procter, P .
CLINICAL BIOMECHANICS, 2004, 19 (05) :519-525
[5]
CARROZZELLA J, 1988, HAND CLIN, V4, P391
[6]
3-DIMENSIONAL FORCE ANALYSIS OF FINGER JOINTS IN SELECTED ISOMETRIC HAND FUNCTIONS [J].
CHAO, EY ;
OPGRANDE, JD ;
AXMEAR, FE .
JOURNAL OF BIOMECHANICS, 1976, 9 (06) :387-396
[7]
Selection of fixation devices in proximal femur rotational osteotomy: clinical complications and finite element analysis [J].
Chen, WP ;
Tai, CL ;
Shih, CH ;
Hsieh, PH ;
Leou, MC ;
Lee, MS .
CLINICAL BIOMECHANICS, 2004, 19 (03) :255-262
[8]
BIOMECHANICAL ANALYSIS OF STATIC FORCES IN THUMB DURING HAND FUNCTION [J].
COONEY, WP ;
CHAO, EYS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (01) :27-36
[9]
Finite element analysis of the mechanical behavior of a scapula implanted with a glenoid prosthesis [J].
Couteau, B ;
Mansat, P ;
Estivalèzes, É ;
Darmana, R ;
Mansat, M ;
Egan, J .
CLINICAL BIOMECHANICS, 2001, 16 (07) :566-575
[10]
Dunning CE, 1999, J HAND SURG-AM, V24A, P992