Ceramic acetabular cups for total hip replacement part 7: How are range of motion and impingement influenced by the position of the centre of rotation and the CCD angle of the stem?

被引:18
作者
Bader, R
Willmann, G
机构
[1] CeramTec AG, D-73207 Plochingen, Germany
[2] FHU Hsch Techn Ulm, Fachbereich Feinwerktechn, Ulm, Germany
来源
BIOMEDIZINISCHE TECHNIK | 1999年 / 44卷 / 12期
关键词
total hip replacement; cup design; ceramic insert; CCD angle; impingement; dislocation;
D O I
10.1515/bmte.1999.44.12.345
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The range of motion (ROM) of total hip prostheses is influenced by a number of parameters. An insufficient ROM may cause impingement, which may result in subluxation, dislocation or material failure of the prostheses. In a three-dimensional CAD simulation, the position of the centre of rotation and the CCD angle of the stem were investigated. Displacement of the centre of rotation of the femoral head may be due to wear (PE cups) or to the design of the prosthesis (ceramic cups). Stems of widely differing design have been developed and implanted. The results of the present study demonstrate that the ROM is clearly reduced by increasing penetration of the femoral head. At an inclination angle of 45 degrees, a depth of penetration of 2 mm restricts flexion by about 15', and a depth of penetration of 3 mm by about 30 degrees. At smaller angles of inclination the ROM is reduced and flexion and abduction are associated with an increased risk of impingement. With steeper acetabular cup inclinations, the risk of impingement decreases, but dislocation, the risk of rim fractures (ceramic cups), and wear and penetration rates (PE cups) increase. The CCD angle of the stent should be oriented to the anatomical situation. At high CCD angles (>135 degrees), flexion is clearly limited, in particular when there is penetration of the femoral head. For modern total hip arthroplasty, prosthetic systems characterised by precise positioning of components, minimum wear, slightly recessed inserts, and appropriate CCD angles should be used.
引用
收藏
页码:345 / 351
页数:7
相关论文
共 18 条
[1]
Ceramic acetabular cups for total hip replacement - Part 6: Cup design, inclination angle and antetorsion angle have an influence on the range of motion and impingement [J].
Bader, R ;
Willmann, G .
BIOMEDIZINISCHE TECHNIK, 1999, 44 (7-8) :212-219
[2]
CHANDLER DR, 1982, CLIN ORTHOP RELAT R, P284
[3]
Effects of acetabular abduction on cup wear rates in total hip arthroplasty [J].
Del Schutte, H ;
Lipman, AJ ;
Bannar, SM ;
Livermore, JT ;
Ilstrup, D ;
Morrey, BF .
JOURNAL OF ARTHROPLASTY, 1998, 13 (06) :621-626
[4]
FROHLING M, 1998, Z ORTHOP, V136, pA55
[5]
FROHLING M, 1995, HUFTENDOPROTHESE KOM, P13
[6]
DISLOCATIONS AFTER TOTAL HIP-ARTHROPLASTY [J].
GROSSMANN, P ;
BRAUN, M ;
BECKER, W .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1994, 132 (06) :521-526
[7]
Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration [J].
Kennedy, JG ;
Rogers, WB ;
Soffe, KE ;
Sullivan, RJ ;
Griffen, DG ;
Sheehan, LJ .
JOURNAL OF ARTHROPLASTY, 1998, 13 (05) :530-534
[8]
Krismer M, 1997, ORTHOPADE, V26, P229
[9]
MORSCHER E, 1995, ENDOPROTHETIK, P157
[10]
NEVELOS J, 1999, 9 C EUR ORTH RES SOC