Electrical failure during cardiopulmonary bypass: an evaluation of incidence, causes, management and guidelines for preventative measures

被引:9
作者
Hargrove, M [1 ]
Ramish, BC [1 ]
O'Donnell, A [1 ]
Aherne, T [1 ]
机构
[1] Cork Univ Hosp, Cardiothorac Surg Unit, Cork, Ireland
来源
PERFUSION-UK | 2002年 / 17卷 / 05期
关键词
D O I
10.1191/0267659102pf592oa
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The incidence of electrical failure during cardiopulmonary bypass (CPB) has been reported to occur in approximately 1 per 1000 cases.(1,2) While the resultant morbidity and mortality is low, electrical failure is a life-threatening scenario. We report three major electrical failures during CPB in a patient population of 3500 over a 15-year period. These cases involved mains failure and generator shut down, mains failure and generator power surge, and failure of the uninterruptable power supply (UPS), which caused protected sockets to shut down. Protocols for preventative maintenance, necessary equipment, battery backup and guidelines for the successful management of such accidents during CPB are discussed.
引用
收藏
页码:369 / 372
页数:4
相关论文
共 7 条
[1]  
HARGROVE M, 1989, J EXTRACORPOR TECHNO, V21, P28
[2]  
KURUSZ M, 1986, P AM ACAD CARDIOVASC, V7, P57
[3]  
NIXON MC, 1997, ANAESTHESIA, V52, P84
[4]   An update on perfusion safety: does the type of perfusion practice affect the rate of incidents related to cardiopulmonary bypass? [J].
Stammers, AH ;
Mejak, BL .
PERFUSION-UK, 2001, 16 (03) :189-198
[5]   AIR-EMBOLISM AND OTHER ACCIDENTS USING PUMP OXYGENATORS [J].
STONEY, WS ;
ALFORD, WC ;
BURRUS, GR ;
GLASSFORD, DM ;
THOMAS, CS .
ANNALS OF THORACIC SURGERY, 1980, 29 (04) :336-340
[6]   COMPLETE ELECTRICAL FAILURE DURING CARDIOPULMONARY BYPASS [J].
TROIANOS, CA .
ANESTHESIOLOGY, 1995, 82 (01) :298-302
[7]  
Welch R H, 1989, J Clin Anesth, V1, P358, DOI 10.1016/0952-8180(89)90076-7