Temporal feature estimation during walking using miniature accelerometers: an analysis of gait improvement after hip arthroplasty

被引:118
作者
Aminian, K [1 ]
Rezakhanlou, K
De Andres, E
Fritsch, C
Leyvraz, PF
Robert, P
机构
[1] Swiss Fed Inst Technol, Metrol Lab, CH-1015 Lausanne, Switzerland
[2] Hop Orthoped Suisse Romande, Lausanne, Switzerland
关键词
accelerometry; gait analysis; orthopaedic; portable device;
D O I
10.1007/BF02513368
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A new method for the detection of gait cycle phases using only two miniature accelerometers together with a light, portable digital recorder is proposed. Each subject is asked to walk on a walkway at his/her own preferred speed. Gait analysis was performed using an original method of computing the values of temporal parameters from accelerometer signals. First, to validate the accelerometric method, measurements are taken on a group of healthy subjects. No significant differences are observed between the results thus obtained and those from pressure sensors attached under the foot Then, measurements using only accelerometers are performed on a group of 12 patients with unilateral hip osteo-arthritis. The gait analysis is carried out just before hip arthroplasty and again, three, six and nine months afterwards. A mean decrease of 88% of asymmetry of stance time and especially a mean decrease of 250% of asymmetry of double support time are observed, nine months after the operation. These results confirm the validity of the proposed method for healthy subjects and its efficiency for functional evaluation of gait improvement after arthroplasty.
引用
收藏
页码:686 / 691
页数:6
相关论文
共 21 条
[1]  
AbuFaraj ZO, 1997, J REHABIL RES DEV, V34, P187
[2]  
AMINIAN K, 1995, MED SCI SPORT EXER, V27, P226
[3]   ESTIMATION OF SPEED AND INCLINE OF WALKING USING NEURAL-NETWORK [J].
AMINIAN, K ;
ROBERT, P ;
JEQUIER, E ;
SCHUTZ, Y .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1995, 44 (03) :743-746
[4]  
Aminian K, 1998, LECT NOTES ARTIF INT, V1537, P1
[5]  
Andersson G, 1972, J Bone Joint Surg Br, V54, P621
[6]   Prediction of human gait parameters from temporal measures of foot-ground contact [J].
Breit, GA ;
Whalen, RT .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1997, 29 (04) :540-547
[7]  
CHAN RB, 1991, ADV BIOENG, V20, P525
[8]   RECORDING ACCELERATIONS IN BODY MOVEMENTS [J].
EVANS, AL ;
DUNCAN, G ;
GILCHRIST, W .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1991, 29 (01) :102-104
[9]   ELITE - A DIGITAL DEDICATED HARDWARE SYSTEM FOR MOVEMENT ANALYSIS VIA REAL-TIME TV SIGNAL-PROCESSING [J].
FERRIGNO, G ;
PEDOTTI, A .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1985, 32 (11) :943-950
[10]  
GIANNINI S, 1994, GAIT ANAL METHODOLGI