Determination of artemisinin in human serum by high-performance liquid chromatography with on-line UV irradiation and peroxyoxalate chemiluminescence detection

被引:21
作者
Amponsaa-Karikari, Abena [1 ]
Kishikawa, Naoya [1 ]
Ohba, Yoshihito [1 ]
Nakashima, Kenichiro [1 ]
Kuroda, Naotaka [1 ]
机构
[1] Nagasaki Univ, Grad Sch Biomed Sci, Nagasaki 8528521, Japan
关键词
arteminsinin; internal endoperoxide linkage; ultraviolet irradiation; high-performance liquid chromatography; peroxyoxalate chemiluminescence;
D O I
10.1002/bmc.664
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Artemisinin is an antimalarial drug containing an internal endoperoxide linkage in its structure. A simple, selective and sensitive high-performance liquid chromatography (HPLC)-peroxyoxalate chemiluminescence (PO-CL) method for the determination of artemisinin was developed. This method is based on the fact that endoperoxide in artemisinin structure can be converted to hydrogen peroxide (H-2,O-2) under ultraviolet (UV) irradiation and the generated hydrogen peroxide can be measured using PO-CL detection. The HPLC-PO-CL system was optimized on a mobile phase, post column chemiluminescence reagent, UV source and irradiation time. In addition, the system was combined with simple liquid-liquid extraction using n-hexane that allowed selective and sensitive determination of artemisinin in serum. The limit of detection using 0.5 mL of blood was 0.062 mu mol/L (17.5 ng/mL) at a signal-to-noise ratio of 3. Calibration curve obtained for artemisinin in human serum 4-80 mu mol/L (1.1-22.6 mu g/mL) showed a good linearity (r = 0.999). Copyright (c) 2006 John Wiley & Sons, Ltd.
引用
收藏
页码:1157 / 1162
页数:6
相关论文
共 27 条
[1]  
Ashton M, 1998, BIOPHARM DRUG DISPOS, V19, P245, DOI 10.1002/(SICI)1099-081X(199805)19:4<245::AID-BDD99>3.3.CO
[2]  
2-Q
[3]   Rapid determination of artemisinin and related analogues using high-performance liquid chromatography and an evaporative light scattering detector [J].
Avery, BA ;
Venkatesh, KK ;
Avery, MA .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 1999, 730 (01) :71-80
[4]  
Dhingra V, 1999, LIFE SCI, V66, P279, DOI 10.1016/S0024-3205(99)00356-2
[5]  
Dwight L., 1995, J CHROMATOGR B, V666, P183
[6]  
EDLUND PO, 1984, ACTA PHARM SUEC, V21, P223
[7]   ANALYSIS OF THE ANTIMALARIAL SESQUITERPENE ARTEMISININ IN ARTEMISIA ANNUA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY (HPLC) WITH POSTCOLUMN DERIVATIZATION AND ULTRAVIOLET DETECTION [J].
ELSOHLY, HN ;
CROOM, EM ;
ELSOHLY, MA .
PHARMACEUTICAL RESEARCH, 1987, 4 (03) :258-260
[8]   A COMPARISON OF GAS-CHROMATOGRAPHY AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY FOR ARTEMISININ ANALYSES [J].
FERREIRA, JFS ;
CHARLES, DJ ;
WOOD, K ;
JANICK, J ;
SIMON, JE .
PHYTOCHEMICAL ANALYSIS, 1994, 5 (03) :116-120
[9]   Direct analysis of artemisinin in plasma and saliva using coupled-column high-performance liquid chromatography with a restricted-access material pre-column [J].
Gordi, T ;
Nielsen, E ;
Yu, ZX ;
Westerlund, D ;
Ashton, M .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2000, 742 (01) :155-162
[10]  
GORDI T, 2001, ACTA U UPPSALA, V250, P2