Pocket-Sized Transthoracic Echocardiography Device for the Measurement of Cardiac Chamber Size and Function

被引:58
作者
Fukuda, Shota [1 ]
Shimada, Kenei
Kawasaki, Toshihiro
Fujimoto, Hiromi
Maeda, Kumiko
Inanami, Hitoshi
Yoshida, Ken
Jissho, Satoshi
Taguchi, Haruyuki
Yoshiyama, Minoru [2 ]
Yoshikawa, Junichi
机构
[1] Osaka Ekisaikai Hosp, Dept Med, Div Cardiovasc, Nishi Ku, Osaka 5500022, Japan
[2] Osaka City Univ, Sch Med, Dept Internal Med & Cardiol, Osaka 545, Japan
关键词
Echocardiography; Imaging; Ventricular function; PERSONAL ULTRASOUND IMAGER; HAND-CARRIED ULTRASOUND; HEART-FAILURE; STANDARDS COMMITTEE; AMERICAN-SOCIETY; CLINICAL UTILITY; RECOMMENDATIONS; HYPERTENSION; FEASIBILITY; DYSFUNCTION;
D O I
10.1253/circj.CJ-08-1076
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: A pocket-sized portable transthoracic echocardiographic (pTTE) imaging device is commercially available, but its feasibility and accuracy in the assessment of cardiac chamber size and function has not been fully compared with the results of standard TTE (sTTE) examination. Methods and Results: The target population comprised 125 unselected patients who underwent sTTE and pTTE examinations. The left ventricular (LV) diastolic and systolic dimensions, fractional shortening (FS), the thickness of the interventricular septum (IVS) and of the LV posterior wall (PWT), left atrial (LA) dimension, and ascending aorta diameter were measured. Echocardiographic measurements were completed for both pTTE and sTTE in all patients (feasibility 100%). LV dimensions, FS, IVS, PWT, LA dimension, and aorta diameter obtained by pTTE showed excellent correlation and agreement with sTTE (r=0.87-0.98, all P<0.001). Observer variabilities for these measurements were similar between pTTE and sTTE. Conclusions: In the present study, pTTE with the Acuson P10 was feasible and accurate for assessing cardiac chamber size and function. (Circ J 2009; 73: 1092-1096)
引用
收藏
页码:1092 / 1096
页数:5
相关论文
共 21 条
[1]   Can the clinical examination diagnose left-sided heart failure in adults? [J].
Badgett, RG ;
Lucey, CR ;
Mulrow, CD .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1997, 277 (21) :1712-1719
[2]   THE LIMITED ECHOCARDIOGRAM - A MODIFICATION OF STANDARD ECHOCARDIOGRAPHY FOR USE IN THE ROUTINE EVALUATION OF PATIENTS WITH SYSTEMIC HYPERTENSION [J].
BLACK, HR ;
WELTIN, G ;
JAFFE, CC .
AMERICAN JOURNAL OF CARDIOLOGY, 1991, 67 (11) :1027-1030
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Personal ultrasound imager: Abdominal aortic aneurysm screening [J].
Bruce, CJ ;
Spittell, PC ;
Montgomery, SC ;
Bailey, KR ;
Tajik, AJ ;
Seward, JB .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2000, 13 (07) :674-679
[5]   Normal Values of Echocardiographic Parameters in Relation to Age in a Healthy Japanese Population-The JAMP Study- [J].
Daimon, Masao ;
Watanabe, Hiroyuki ;
Abe, Yukio ;
Hirata, Kumiko ;
Hozumi, Takeshi ;
Ishii, Katsuhisa ;
Ito, Hiroshi ;
Iwakura, Katsuomi ;
Izumi, Chisato ;
Matsuzaki, Masunori ;
Minagoe, Shinichi ;
Abe, Haruhiko ;
Murata, Kazuya ;
Nakatani, Satoshi ;
Negishi, Kazuaki ;
Yoshida, Ken ;
Tanabe, Kazuaki ;
Tanaka, Nobuhiro ;
Tokai, Kotaro ;
Yoshikawa, Junichi .
CIRCULATION JOURNAL, 2008, 72 (11) :1859-1866
[6]   The pocket echocardiograph: a useful new tool? [J].
Egan, Margaret ;
Ionescu, A. .
EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY, 2008, 9 (06) :721-725
[7]   Utility of hand-carried ultrasound for consultative cardiology [J].
Gorcsan, J .
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2003, 20 (05) :463-469
[8]   The use of hand-carried ultrasound in the hospital setting - A cost-effective analysis [J].
Greaves, K ;
Jeetley, P ;
Hickman, M ;
Dwivedi, G ;
Sabharwal, N ;
Lim, T ;
Janardhanan, R ;
Senior, R .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2005, 18 (06) :620-625
[9]   Feasibility of "limited" echo imaging: Characterization of incidental findings [J].
Kimura, BJ ;
Pezeshki, B ;
Frack, SA ;
DeMaria, AN .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 1998, 11 (07) :746-750
[10]   Recommendations for chamber quantification: A report from the American Society of Echocardiography's guidelines and standards committee and the chamber quantification writing group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology [J].
Lang, RM ;
Bierig, M ;
Devereux, RB ;
Flachskampf, FA ;
Foster, E ;
Pellikka, PA ;
Picard, MH ;
Roman, MJ ;
Seward, J ;
Shanewise, JS ;
Solomon, SD ;
Spencer, KT ;
Sutton, MS ;
Stewart, WJ .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2005, 18 (12) :1440-1463